Impact of respiratory motion correction and spatial resolution on lesion detection in PET: a simulation study based on real MR dynamic data
Authors:
- Irene PolycarpouIPIrene Polycarpou
- Charalampos TsoumpasCTCharalampos Tsoumpas
- Andrew P KingAKAndrew P King
- Paul K MarsdenPMPaul K Marsden
Open Access
- 20 January 2014
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (3) , 697-713
Abstract
The aim of this study is to investigate the impact of respiratory motion correction and spatial resolution on lesion detectability in PET as a function of lesion size and tracer uptake. Real respiratory signals describing different breathing types are combined with a motion model formed from real dynamic MR data to simulate multiple dynamic PET datasets acquired from a continuously moving subject. Lung and liver lesions were simulated with diameters ranging from 6 to 12 mm and lesion to background ratio ranging from 3:1 to 6:1. Projection data for 6 and 3 mm PET scanner resolution were generated using analytic simulations and reconstructed without and with motion correction. Motion correction was achieved using motion compensated image reconstruction. The detectability performance was quantified by a receiver operating characteristic (ROC) analysis obtained using a channelized Hotelling observer and the area under the ROC curve (AUC) was calculated as the figure of merit. The results indicate that respiratory motion limits the detectability of lung and liver lesions, depending on the variation of the breathing cycle length and amplitude. Patients with large quiescent periods had a greater AUC than patients with regular breathing cycles and patients with long-term variability in respiratory cycle or higher motion amplitude. In addition, small (less than 10 mm diameter) or low contrast (3:1) lesions showed the greatest improvement in AUC as a result of applying motion correction. In particular, after applying motion correction the AUC is improved by up to 42% with current PET resolution (i.e. 6 mm) and up to 51% for higher PET resolution (i.e. 3 mm). Finally, the benefit of increasing the scanner resolution is small unless motion correction is applied. This investigation indicates high impact of respiratory motion correction on lesion detectability in PET and highlights the importance of motion correction in order to benefit from the increased resolution of future PET scanners.Keywords
This publication has 48 references indexed in Scilit:
- The effect of regularization in motion compensated PET image reconstruction: a realistic numerical 4D simulation studyPhysics in Medicine & Biology, 2013
- Respiratory motion models: A reviewMedical Image Analysis, 2013
- Analysis and comparison of two methods for motion correction in PET imagingMedical Physics, 2012
- MRI-Based Nonrigid Motion Correction in Simultaneous PET/MRIJournal of Nuclear Medicine, 2012
- Introducing Parametric Fusion PET/MRI of Primary Prostate CancerJournal of Nuclear Medicine, 2012
- Fast generation of 4D PET-MR data from real dynamic MR acquisitionsPhysics in Medicine & Biology, 2011
- Hierarchical adaptive local affine registration for fast and robust respiratory motion estimationMedical Image Analysis, 2011
- Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluationPhysics in Medicine & Biology, 2010
- The impact of respiratory motion on tumor quantification and delineation in static PET/CT imagingPhysics in Medicine & Biology, 2009
- Variability of Four-Dimensional Computed Tomography Patient ModelsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008