A fast double template convolution isocenter evaluation algorithm with subpixel accuracy
Authors:
- Brian WineyBWBrian Winey
- Greg SharpGSGreg Sharp
- Marc BussièreMBMarc Bussière
- 20 December 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 38 (1) , 223-227
Abstract
To design a fast Winston Lutz (fWL) algorithm for accurate analysis of radiation isocenter from images without edge detection or center of mass calculations. An algorithm has been developed to implement the Winston Lutz test for mechanical/ radiation isocenter agreement using an electronic portal imaging device (EPID). The algorithm detects the position of the radiation shadow of a tungsten ball within a stereotactic cone. The fWL algorithm employs a double convolution to independently find the position of the sphere and cone centers. Subpixel estimation is used to achieve high accuracy. Results of the algorithm were compared to (1) a human observer with template guidance and (2) an edge detection/center of mass (edCOM) algorithm. Testing was performed with high resolution (0.05 mm/px, film) and low resolution (0.78 mm/px, EPID) image sets. Sphere and cone center relative positions were calculated with the fWL algorithm for high resolution test images with an accuracy of 0.002 +/- 0.061 mm compared to 0.042 +/- 0.294 mm for the human observer, and 0.003 +/- 0.038 mm for the edCOM algorithm. The fWL algorithm required 0.01 s per image compared to 5 s for the edCOM algorithm and 20 s for the human observer. For lower resolution images the fWL algorithm localized the centers with an accuracy of 0.083 +/- 0.12 mm compared to 0.03 +/- 0.5514 mm for the edCOM algorithm. A fast (subsecond) subpixel algorithm has been developed that can accurately determine the center locations of the ball and cone in Winston Lutz test images without edge detection or COM calculations.Keywords
This publication has 5 references indexed in Scilit:
- A robust Hough transform algorithm for determining the radiation centers of circular and rectangular fields with subpixel accuracyPhysics in Medicine & Biology, 2009
- Efficient subpixel image registration algorithmsOptics Letters, 2008
- Introducing a system for automated control of rotation axes, collimator and laser adjustment for a medical linear acceleratorPhysics in Medicine & Biology, 2003
- Verification of radiosurgery target point alignment with an electronic portal imaging device (EPID)Medical Physics, 1997
- Linear Accelerator as a Neurosurgical Tool for Stereotactic RadiosurgeryNeurosurgery, 1988