Robust incremental compensation of the light attenuation with depth in 3D fluorescence microscopy
- 21 May 2004
- journal article
- Published by Wiley in Journal of Microscopy
- Vol. 214 (3) , 297-314
Abstract
Summary: Fluorescent signal intensities from confocal laser scanning microscopes (CLSM) suffer from several distortions inherent to the method. Namely, layers which lie deeper within the specimen are relatively dark due to absorption and scattering of both excitation and fluorescent light, photobleaching and/or other factors. Because of these effects, a quantitative analysis of images is not always possible without correction. Under certain assumptions, the decay of intensities can be estimated and used for a partial depth intensity correction. In this paper we propose an original robust incremental method for compensating the attenuation of intensity signals. Most previous correction methods are more or less empirical and based on fitting a decreasing parametric function to the section mean intensity curve computed by summing all pixel values in each section. The fitted curve is then used for the calculation of correction factors for each section and a new compensated sections series is computed. However, these methods do not perfectly correct the images. Hence, the algorithm we propose for the automatic correction of intensities relies on robust estimation, which automatically ignores pixels where measurements deviate from the decay model. It is based on techniques adopted from the computer vision literature for image motion estimation. The resulting algorithm is used to correct volumes acquired in CLSM. An implementation of such a restoration filter is discussed and examples of successful restorations are given.Keywords
This publication has 27 references indexed in Scilit:
- Robust multiresolution alignment of MRI brain volumesMagnetic Resonance in Medicine, 2000
- Heterogeneous photobleaching in confocal microscopy caused by differences in refractive index and excitation modeCytometry, 1998
- Three-dimensional DNA image cytometry by confocal scanning laser microscopy in thick tissue blocks of prostatic lesionsCytometry, 1997
- The Robust Estimation of Multiple Motions: Parametric and Piecewise-Smooth Flow FieldsComputer Vision and Image Understanding, 1996
- Robust Multiresolution Estimation of Parametric Motion ModelsJournal of Visual Communication and Image Representation, 1995
- Fluorophores for Confocal Microscopy1995
- The Collection, Processing, and Display of Digital Three-Dimensional Images of Biological Specimens1995
- Handbook of Biological Confocal Microscopy1990
- Robust Regression and Outlier Detection1987
- Robust Statistics1981