A plane-grating monochromator for radiometric calibration
Authors:
- 1 July 1989
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 60 (7) , 2089-2092
Abstract
A new plane-grating monochromator and calibration chamber were constructed for the radiometric calibration of plasma diagnostic devices. The monochromator is designed to cover a wavelength range of 1.8–238 nm with moderate resolving power of 500±200 in λ/Δλ. To check the performance of the monochromator, spectral distribution and resolution were measured by experiments on photoemission from solids and photoionization of rare gases. The characterization of multilayer reflectors by the use of this beamline is demonstrated.Keywords
This publication has 10 references indexed in Scilit:
- New far UV detector calibration facility at the National Bureau of StandardsApplied Optics, 1987
- A plane-grating monochromator for 2 eV ≤ hv ≤ 150 eVNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1986
- Curved Crystal Spectrometer with Position-Sensitive Solid-State Detector for Plasma DiagnosticsJapanese Journal of Applied Physics, 1983
- The use of synchrotron radiation for detector calibrationsNuclear Instruments and Methods, 1980
- The use of synchrotron radiation as an absolute source of VUV radiationJournal of Research of the National Bureau of Standards Section A: Physics and Chemistry, 1975
- Absolute Radiometric Calibration of Detectors Between 200–600 ÅApplied Optics, 1975
- Soft X-Ray Absorption Spectra of Alkali Halides. I. KCl and NaClJournal of the Physical Society of Japan, 1966
- Absolute Intensity Measurements in the Vacuum Ultraviolet*†Journal of the Optical Society of America, 1964
- Absorption Coefficients of Xenon and Argon in the 600–1025 Å Wavelength RegionsThe Journal of Chemical Physics, 1963
- Tracing Skew Rays through Concave Diffraction GratingsOptica Acta: International Journal of Optics, 1962