Some Aspects of Nonstationary Myoelectric Signal Processing
Authors:
- 1 February 1987
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-34 (2) , 166-172
Abstract
Nonstationary myoelectric signal processing is considered. A midpoint moving average estimator for the signal variance is developed. The performance of the estimator is analyzed and optimum window size criteria are derived. Comparison of the theoretical performance to experimental performance shows good agreement. The resultant estimation error is only moderately (2-3 percent) degraded from the equivalent stationary case. Finally, the developed estimator is compared to a typical squarer followed by a low-pass filter myoelectric processor. Provided that the low-pass filter has a linear phase characteristic in the passband, the two systems have essentially the same error performance.Keywords
This publication has 11 references indexed in Scilit:
- Frequency Parameters of the Myoelectric Signal as a Measure of Muscle Conduction VelocityIEEE Transactions on Biomedical Engineering, 1981
- Effects of Load on Myoelectric Signals: The ARIMA RepresentationIEEE Transactions on Biomedical Engineering, 1981
- Myoelectric Signal Processing: Optimal Estimation Applied to Electromyography - Part I: Derivation of the Optimal MyoprocessorIEEE Transactions on Biomedical Engineering, 1980
- Sequential Multistate EMG Signal ProcessorIEEE Transactions on Biomedical Engineering, 1979
- Physiology and Mathematics of Myoelectric SignalsIEEE Transactions on Biomedical Engineering, 1979
- A Nonstationary Model for the ElectromyogramIEEE Transactions on Biomedical Engineering, 1977
- Signal processing for the multistate myoelectric channelProceedings of the IEEE, 1977
- Functional Separation of EMG Signals via ARMA Identification Methods for Prosthesis Control PurposesIEEE Transactions on Systems, Man, and Cybernetics, 1975
- Dynamic relationship between isometric muscle tension and the electromyogram in man.Journal of Applied Physiology, 1971
- Signal Versus Noise Characteristics of Filtered EMG Used as a Control SourceIEEE Transactions on Biomedical Engineering, 1971