Rotary Blood Pump Control Strategy for Preventing Left Ventricular Suction
- 1 January 2015
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 61 (1) , 21-30
Abstract
The risk for left ventricular (LV) suction while maintaining adequate perfusion over a range of physiologic conditions during continuous flow LV assist device (LVAD) support is a significant clinical concern. To address this challenge, we developed a suction prevention and physiologic control (SPPC) algorithm for use with axial and centrifugal LVADs. The SPPC algorithm uses two gain-scheduled, proportional-integral controllers that maintain a differential pump speed (ΔRPM) above a user-defined threshold to prevent LV suction, while maintaining an average reference differential pressure (ΔP) between the LV and aorta to provide physiologic perfusion. Efficacy and robustness of the proposed algorithm were evaluated in silico during simulated rest and exercise test conditions for (1) ΔP/ΔRPM excessive setpoint (ES); (2) rapid eightfold increase in pulmonary vascular resistance (PVR); and (3) ES and PVR. Hemodynamic waveforms (LV pressure and volume; aortic pressure and flow) were simulated and analyzed to identify suction event(s), quantify total flow output (pump + cardiac output), and characterize the performance of the SPPC algorithm. The results demonstrated that the proposed SPPC algorithm prevented LV suction while maintaining physiologic perfusion for all simulated test conditions, and warrants further investigation in vivo.Keywords
This publication has 28 references indexed in Scilit:
- A Suction Detection System for Rotary Blood Pumps Based on the Lagrangian Support Vector Machine AlgorithmIEEE Journal of Biomedical and Health Informatics, 2013
- Response of Rotary Blood Pumps to Changes in Preload and Afterload at a Fixed Speed Setting Are Unphysiological When Compared With the Natural HeartArtificial Organs, 2011
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceThe New England journal of medicine, 2009
- NONINVASIVE DETECTION OF SUCTION IN AN IMPLANTABLE ROTARY BLOOD PUMP USING NEURAL NETWORKSInternational Journal of Computational Intelligence and Applications, 2008
- Advanced Suction Detection for an Axial Flow PumpArtificial Organs, 2006
- Identification and Classification of Physiologically Significant Pumping States in an Implantable Rotary Blood PumpArtificial Organs, 2006
- A discriminant-analysis-based suction detection system for rotary blood pumps2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2006
- Development of a Suction Detection System for Axial Blood PumpsArtificial Organs, 2004
- Sixth International Symposium for Rotary Blood PumpsArtificial Organs, 1999
- Detection of Suction and Regurgitation of the Implantable Centrifugal Pump Based on the Motor Current Waveform Analysis and Its Application to Optimization of Pump FlowArtificial Organs, 1999