Spectral velocity profiles for detailed ultrasound flow analysis
- 1 July 1996
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 43 (4) , 654-659
Abstract
The operation of a novel ultrasound multigate instrument capable of computing in real-time the fast Fourier transform (FFT) of Doppler signals detected from 64 equally spaced range cells is presented. The new system provides up to 50 velocity profiles per second, which are displayed in such a manner that information about the full spectral content of Doppler signals at all the investigated depths is continuously monitored over a PRF-wide frequency range which can be set arbitrarily between -PRF and +PRF. Experimental results are presented, which demonstrate that the true velocity profile can be accurately detected through the computation of "local" maximum velocities obtained by properly correcting the maximum frequency of each spectrum. There is also a discussion on how the results of multigate analysis are influenced by the sample volume length, a parameter which can be usually set by modifying the duration of the transmitted burst. In particular, it is shown that, in regions close to the vessel walls, the shear rate can be measured with a spatial resolution related to the spacing between subsequent range cells and not to the sample volume length.Keywords
This publication has 17 references indexed in Scilit:
- A real-time ultrasound time-domain correlation blood flowmeter. II. Performance and experimental verificationIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1993
- Experimental evaluation of intrinsic and nonstationary ultrasonic Doppler spectral broadening in steady and pulsatile flow loop modelsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1993
- Measurement of mean velocity during pulsatile flow using time-averaged maximum frequency of doppler ultrasound waveformsUltrasound in Medicine & Biology, 1993
- Spectral composition of doppler signalsUltrasound in Medicine & Biology, 1991
- Time Domain Formulation of Pulse-Doppler Ultrasound and Blood Velocity Estimation by Cross CorrelationUltrasonic Imaging, 1986
- Velocity Profile Reconstruction Using Ultrafast Spectral Analysis of Doppler UltrasoundIEEE Transactions on Sonics and Ultrasonics, 1985
- The Accurate Ultrasonic Measurement of the Volume Flow of Blood by Time Domain Correlation1985
- 25 + 1 channel pulsed ultrasound Doppler velocity meter for quantitative flow measurements and turbulence analysisUltrasound in Medicine & Biology, 1984
- Limitations of doppler spectral broadening in the early detection of carotid artery disease due to the size of the sample volumeUltrasound in Medicine & Biology, 1983
- Topoflow - A digital full range Doppler velocity meterIEEE Transactions on Sonics and Ultrasonics, 1978