Sensitive detection of cardiac troponin T based on superparamagnetic bead-labels using a flexible micro-fluxgate sensor
Open Access
- 10 November 2017
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in RSC Advances
- Vol. 7 (82) , 52327-52336
Abstract
In this study, we describe an innovative micro-fluxgate immunosensor based on superparamagnetic Dynabeads for the detection of cardiac troponin T (cTnT), an important biomarker for cardiovascular diseases. The fluxgate sensor developed in this study uses a double layer of Co-based amorphous ribbons as a sensing element and three-dimensional solenoid copper coils to control the sensitive core. The system was fabricated through a standard micro-fabrication process, including thick photoresist lithography and electroplating. Superparamagnetic Dynabeads were employed as recognition tags and double polyclonal antibody pairs combined using streptavidin–biotin binding were employed to specifically capture and label the cTnT coupled with magnetic beads on a separate Au film-coated wafer via a classical sandwich immunoassay process. Series optimization of the assay parameters was performed to achieve optimal detection parameters. Under optimal conditions, the detection and quantification of cTnT was performed by sensing the magnetic signal of the cTnT-labeled Dynabeads with the micro-fluxgate sensor. The resulting biosystem successfully detected cTnT with satisfactory sensitivity, reproducibility, stability, and specificity. A minimum detectable limit of 0.01 ng mL−1 was achieved with a linearity range of 0.01–10 ng mL−1, demonstrating the high sensitivity of this novel system. In addition to the superior detection performance, the proposed sensor is convenient to manipulate, entirely lab-free, provides a quick result, and is portable. These factors suggest that this system has the potential to be applied as a point-of-care diagnostic tool for cardiovascular disease.Keywords
Funding Information
- Natural Science Foundation of Shanghai Municipality (13ZR1420800)
- Natural Science Foundation of Henan Province (162300410233)
- National Natural Science Foundation of China (61273065)
- Shanghai Jiao Tong University (AgriX2015005)
- Shanghai Academy of Spaceflight Technology (USCAST2015-2)
This publication has 30 references indexed in Scilit:
- Signal dependence on magnetic nanoparticles position over a planar Hall effect biosensorMicroelectronic Engineering, 2013
- An o-aminobenzoic acid film-based immunoelectrode for detection of the cardiac troponin T in human serumBiochemical Engineering Journal, 2013
- Improved micro fluxgate sensor with double-layer Fe-based amorphous coreMicrosystem Technologies, 2012
- Fabrication and characterization of a new MEMS fluxgate sensor with nanocrystalline magnetic coreMeasurement, 2012
- A dual quartz crystal microbalance for human cardiac troponin T in real time detectionSensors and Actuators B: Chemical, 2012
- A Nanostructured Piezoelectric Immunosensor for Detection of Human Cardiac Troponin TSensors, 2011
- Cardiac troponin T detection using polymers coated quartz crystal microbalance as a cost-effective immunosensorBiomedical Engineering / Biomedizinische Technik, 2010
- Surface plasmon resonance immunosensor for human cardiac troponin T based on self-assembled monolayerJournal of Pharmaceutical and Biomedical Analysis, 2007
- An SPR immunosensor for human cardiac troponin T using specific binding avidin to biotin at carboxymethyldextran-modified gold chipClinica Chimica Acta, 2007
- Release of cardiac troponin in acute coronary syndromes: ischemia or necrosis?Clinica Chimica Acta, 1999