Abstract
A sensitive method for simultaneous determination of azithromycin (AZI), acetylspiramycin (ACE), erythromycin (ERY), and josamycin (JOS) was developed by CE coupled with electrochemiluminescence detection with Ru(bpy)32+. The parameters related to separation and detection were investigated in detail. The four macrolides were well separated and detected within 6 min under the optimized conditions. The LOD (S/N=3) of AZI, ACE, ERY, and JOS were 1.2×10−9, 7.1×10−9, 3.9×10−8 and 9.5×10−8 mol/L, respectively. The LOQ (S/N=10) of AZI, ACE, ERY, and JOS in human urine were 8.2×10−8, 2.5×10−7, 8.9×10−7 and 1.2×10−6 mol/L, respectively. The recoveries of the four macrolides in human urine and pharmaceutical tablet samples were 85.0–104.0% at different concentration levels.

This publication has 37 references indexed in Scilit: