Are Ultrathin Stents Optimal for Bifurcation Lesions? Insights From Computational Modeling of Provisional and DK‐Crush Techniques
- 12 November 2025
- journal article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 107 (1) , 258-272
Abstract
Background: Complex coronary bifurcation lesions remain challenging in percutaneous coronary intervention, with stent design and deployment strategy influencing clinical outcomes. Aims: This study compares the mechanical and hemodynamic performance of the ultrathin‐strut Orsiro and thin‐strut Xience Sierra stent in provisional side branch (PSB) and double kissing crush (DKC) techniques. Methods: We used finite element analyses of bifurcation stent deployment to assess malapposition, ostium clearance, and arterial wall stress for both techniques. Computational fluid dynamics simulations quantified the luminal exposure to low time‐averaged endothelial shear stress (TAESS < 0.4 Pa) and high shear rates (> 1000 s⁻¹). Results: In PSB, Orsiro showed higher malapposition (13.0% vs. 9.6%) but improved SB ostium clearance (77% vs. 64%) and lower low‐TAESS exposure (30.3% vs. 33.6%) compared to Xience. Orsiro also produced higher arterial wall stresses, particularly during kissing balloon inflation. In DKC, differences in malapposition and ostium clearance diminished between stents, though Orsiro retained a hemodynamic advantage with lower low‐TAESS (28.2% vs. 36.3%). Conclusions: Stent design influenced outcomes more strongly in PSB, where anatomical interaction and platform‐specific behavior impacted both structural and hemodynamic results. In DKC, procedural complexity minimized those differences, making the stenting technique the primary performance driver. Nonetheless, Orsiro consistently preserved more favorable flow conditions. These findings highlight the need to match device selection with lesion characteristics in PSB, while in DKC, optimizing procedural steps may have a greater impact than the choice of stent platform.Keywords
All Related Versions
This publication has 58 references indexed in Scilit:
- Advancements in Coronary Bifurcation Stenting Techniques: Insights From Computational and Bench Testing StudiesInternational Journal for Numerical Methods in Biomedical Engineering, 2025
- European Bifurcation Club white paper on stenting techniques for patients with bifurcated coronary artery lesionsCatheterization and Cardiovascular Interventions, 2020
- Three contemporary thin‐strut drug‐eluting stents implanted in severely calcified coronary lesions of participants in a randomized all‐comers trialCatheterization and Cardiovascular Interventions, 2020
- Outcomes in Patients Treated With Thin-Strut, Very Thin-Strut, or Ultrathin-Strut Drug-Eluting Stents in Small Coronary VesselsJAMA Cardiology, 2019
- Differences in rotational positioning and subsequent distal main branch rewiring of the Tryton stent: An optical coherence tomography and computational studyCatheterization and Cardiovascular Interventions, 2018
- A Study of Coronary Bifurcation Shape in a Normal PopulationJournal of Cardiovascular Translational Research, 2016
- Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design ConsiderationsAnnals of Biomedical Engineering, 2015
- High Wall Shear Stress and Spatial Gradients in Vascular Pathology: A ReviewAnnals of Biomedical Engineering, 2012
- Modelling of the provisional side-branch stenting approach for the treatment of atherosclerotic coronary bifurcations: effects of stent positioningBiomechanics and Modeling in Mechanobiology, 2010
- Hemodynamic Shear Stress and Its Role in AtherosclerosisJAMA, 1999