Implant Experience with an Implantable Hemodynamic Monitor for the Management of Symptomatic Heart Failure
Authors:
- Dwight W. ReynoldsDRDwight W. Reynolds
- Fredrik GadlerFGFredrik Gadler
- G. Neal KayGKG. Neal Kay
- Mike F. HessMHMike F. Hess
- Tom BennettTBTom Bennett
- For The Chronicle InvestigatorsFIFor The Chronicle Investigators
- 11 August 2005
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 28 (8) , 747-753
Abstract
Introduction: Management of congestive heart failure is a serious public health problem. The use of implantable hemodynamic monitors (IHMs) may assist in this management by providing continuous ambulatory filling pressure status for optimal volume management. Methods and Results: The Chronicle® system includes an implanted monitor, a pressure sensor lead with passive fixation, an external pressure reference (EPR), and data retrieval and viewing components. The tip of the lead is placed near the right ventricular outflow tract to minimize risk of sensor tissue encapsulation. Implant technique and lead placement is similar to that of a permanent pacemaker. After the system had been successfully implanted in 148 patients, the type and frequency of implant‐related adverse events were similar to a single‐chamber pacemaker implant. R‐wave amplitude was 15.2 ± 6.7 mV and the pressure waveform signal was acceptable in all but two patients in whom presence of artifacts required lead repositioning. Implant procedure time was not influenced by experience, remaining constant throughout the study. Conclusion: Based on this evaluation, permanent placement of an IHM in symptomatic heart failure patients is technically feasible. Further investigation is warranted to evaluate the use of the continuous hemodynamic data in management of heart failure patients.Keywords
This publication has 13 references indexed in Scilit:
- Development of Implantable Devices for Continuous Ambulatory Monitoring of Central Hemodynamic Values in Heart Failure PatientsPacing and Clinical Electrophysiology, 2005
- Ongoing right ventricular hemodynamics in heart failureJACC, 2003
- Cardiac Resynchronization in Chronic Heart FailureNew England Journal of Medicine, 2002
- Continuous ambulatory right heart pressure measurements with an implantable hemodynamic monitor: A multicenter, 12-month follow-up study of patients with chronic heart failureJournal of Cardiac Failure, 2002
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionNew England Journal of Medicine, 2002
- Continous haemodynamic monitoring during withdrawal of diuretics in patients with congestive heart failureEuropean Heart Journal, 2002
- Continuous ambulatory monitoring of absolute right ventricular pressure and mixed venous oxygen saturation in patients with heart failure using an implantable haemodynamic monitor: results of a 1 year multicentre feasibility studyEuropean Heart Journal, 2001
- Tailored Therapy to Hemodynamic Goals for Advanced Heart FailureEuropean Journal of Heart Failure, 1999
- Measurement of pulmonary artery diastolic pressure from the right ventricleJACC, 1995
- Monitoring of pulmonary arterial diastolic pressure through a right ventricular pressure transducerJournal of Cardiac Failure, 1995