Adaptive Monostatic System for Measuring Microwave Reflections from the Breast
Open Access
Abstract
A second-generation monostatic radar system to measure microwave reflections from the human breast is presented and analyzed. The present system can measure the outline of the breast with an accuracy of ±1 mm and precisely place the microwave sensor in an adaptive matter such that microwaves are normally incident on the skin. Microwave reflections are measured between 10 MHz to 12 GHz with sensitivity of 65 to 75 dB below the input power and a total scan time of 30 min for 140 locations. The time domain reflections measured from a volunteer show fidelity above 0.98 for signals in a single scan. Finally, multiple scans of a breast phantoms demonstrate the consistency of the system in terms of recorded reflection, outline measurement, and image reconstruction.Keywords
This publication has 18 references indexed in Scilit:
- Novel Microwave Torso Scanner for Thoracic Fluid Accumulation Diagnosis and MonitoringScientific Reports, 2017
- Microwave technology for detecting traumatic intracranial bleedings: tests on phantom of subdural hematoma and numerical simulationsMedical & Biological Engineering & Computing, 2016
- MARIA M4: clinical evaluation of a prototype ultrawideband radar scanner for breast cancer detectionJournal of Medical Imaging, 2016
- Evaluation of 3-D Acquisition Surfaces for Radar-Based Microwave Breast ImagingIEEE Transactions on Antennas and Propagation, 2015
- An Early Clinical Study of Time-Domain Microwave Radar for Breast Health MonitoringIEEE Transactions on Biomedical Engineering, 2015
- Microwave Breast Imaging With a Monostatic Radar-Based System: A Study of Application to PatientsIEEE Transactions on Microwave Theory and Techniques, 2013
- Clinical Microwave Tomographic Imaging of the Calcaneus: A First-in-Human Case Study of Two SubjectsIEEE Transactions on Biomedical Engineering, 2012
- A Prototype System for Measuring Microwave Frequency Reflections from the BreastInternational Journal of Biomedical Imaging, 2012
- Microwave Radar-Based Differential Breast Cancer Imaging: Imaging in Homogeneous Breast Phantoms and Low Contrast ScenariosIEEE Transactions on Antennas and Propagation, 2010
- A large-scale study of the ultrawideband microwave dielectric properties of normal, benign and malignant breast tissues obtained from cancer surgeriesPhysics in Medicine & Biology, 2007