Towards Improved Alignment of Powder Diffractometers
Author:
- 1 March 1986
- journal article
- research article
- Published by Cambridge University Press (CUP) in Powder Diffraction
- Vol. 1 (1) , 26-33
Abstract
An improved method of alignment of a Philips diffractometer equipped with a theta-compensating slit is described. The method employs a special fluorescent screen which is flat to better than 10 microns. The procedure results in accurate alignment of the theta-compensating slit at low angles and assures reliable intensity data down to one degree two theta or less.Keywords