Assessment of an ECG‐Based System for Localizing Ventricular Arrhythmias in Patients With Structural Heart Disease
Open Access
- 19 October 2021
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of the American Heart Association
- Vol. 10 (20) , e022217
Abstract
Background: We have previously developed an intraprocedural automatic arrhythmia‐origin localization (AAOL) system to identify idiopathic ventricular arrhythmia origins in real time using a 3‐lead ECG. The objective was to assess the localization accuracy of ventricular tachycardia (VT) exit and premature ventricular contraction (PVC) origin sites in patients with structural heart disease using the AAOL system. Methods and Results: In retrospective and prospective case series studies, a total of 42 patients who underwent VT/PVC ablation in the setting of structural heart disease were recruited at 2 different centers. The AAOL system combines 120‐ms QRS integrals of 3 leads (III, V2, V6) with pace mapping to predict VT exit/PVC origin site and projects that site onto the patient‐specific electroanatomic mapping surface. VT exit/PVC origin sites were clinically identified by activation mapping and/or pace mapping. The localization error of the VT exit/PVC origin site was assessed by the distance between the clinically identified site and the estimated site. In the retrospective study of 19 patients with structural heart disease, the AAOL system achieved a mean localization accuracy of 6.5±2.6 mm for 25 induced VTs. In the prospective study with 23 patients, mean localization accuracy was 5.9±2.6 mm for 26 VT exit and PVC origin sites. There was no difference in mean localization error in epicardial sites compared with endocardial sites using the AAOL system (6.0 versus 5.8 mm, P =0.895). Conclusions: The AAOL system achieved accurate localization of VT exit/PVC origin sites in patients with structural heart disease; its performance is superior to current systems, and thus, it promises to have potential clinical utility.Keywords
This publication has 20 references indexed in Scilit:
- Real-Time Localization of Ventricular Tachycardia Origin From the 12-Lead ElectrocardiogramJACC: Clinical Electrophysiology, 2017
- Ventricular Tachycardia Ablation versus Escalation of Antiarrhythmic DrugsNew England Journal of Medicine, 2016
- Automated analysis of the 12-lead electrocardiogram to identify the exit site of postinfarction ventricular tachycardiaHeart Rhythm, 2012
- Isolated Potentials During Sinus Rhythm and Pace-Mapping Within Scars as Guides for Ablation of Post-Infarction Ventricular TachycardiaJACC, 2006
- Mapping for Ventricular TachycardiaCardiac Electrophysiology Review, 2002
- Electrically Unexcitable Scar Mapping Based on Pacing Threshold for Identification of the Reentry Circuit IsthmusCirculation, 2002
- Gaining access to the pericardial spaceThe American Journal of Cardiology, 2002
- Radiofrequency Catheter Ablation of Ventricular Tachycardia After Myocardial InfarctionCirculation, 1998
- Exploring Postinfarction Reentrant Ventricular Tachycardia With Entrainment MappingJACC, 1997
- Relationship between the 12-lead electrocardiogram during ventricular tachycardia and endocardial site of origin in patients with coronary artery disease.Circulation, 1988