Long-term follow-up of abandoned transvenous defibrillator leads: a nationwide cohort study
- 23 May 2020
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 22 (7) , 1097-1102
Abstract
Aims : Commonly, a dysfunctional defibrillator lead is abandoned and a new lead is implanted. Long-term follow-up data on abandoned leads are sparse. We aimed to investigate the incidence and reasons for extraction of abandoned defibrillator leads in a nationwide cohort and to describe extraction procedure-related complications. Methods and results : All abandoned transvenous defibrillator leads were identified in the Danish Pacemaker and ICD Register from 1991 to 2019. The event-free survival of abandoned defibrillator leads was studied, and medical records of patients with interventions on abandoned defibrillator leads were audited for procedure-related data. We identified 740 abandoned defibrillator leads. Meantime from implantation to abandonment was 7.2 ± 3.8 years with mean patient age at abandonment of 66.5 ± 13.7 years. During a mean follow-up after abandonment of 4.4 ± 3.1 years, 65 (8.8%) abandoned defibrillator leads were extracted. Most frequent reason for extraction was infection (pocket and systemic) in 41 (63%) patients. Procedural outcome after lead extraction was clinical success in 63 (97%) patients. Minor complications occurred in 3 (5%) patients, and major complications in 1 (2%) patient. No patient died from complication to the procedure during 30-day follow-up after extraction. Conclusion : More than 90% of abandoned defibrillator leads do not need to be extracted during long-term follow-up. The most common indication for extraction is infection. Abandoned defibrillator leads can be extracted with high clinical success rate and low risk of major complications at high-volume centres.Keywords
This publication has 19 references indexed in Scilit:
- Complication Rates Associated With Pacemaker or Implantable Cardioverter-Defibrillator Generator Replacements and Upgrade ProceduresCirculation, 2010
- Extraction of cardiac rhythm devices: indications, techniques and outcomes for the removal of pacemaker and defibrillator leadsInternational Journal Of Clinical Practice, 2010
- Risk of Failure of Transvenous Implantable Cardioverter-Defibrillator LeadsCirculation: Arrhythmia and Electrophysiology, 2009
- Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient ManagementHeart Rhythm, 2009
- Do abandoned leads pose risk to implantable cardioverter-defibrillator patients?Heart Rhythm, 2009
- Necessity for Surgical Revision of Defibrillator Leads Implanted Long-TermCirculation, 2008
- Tutorial in biostatistics: competing risks and multi‐state modelsStatistics in Medicine, 2006
- Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart FailureNew England Journal of Medicine, 2005
- Prophylactic Implantation of a Defibrillator in Patients with Myocardial Infarction and Reduced Ejection FractionNew England Journal of Medicine, 2002
- Improved Survival with an Implanted Defibrillator in Patients with Coronary Disease at High Risk for Ventricular ArrhythmiaNew England Journal of Medicine, 1996