A Fill-and-Flow Biosensor
- 17 June 1998
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 70 (15) , 3131-3136
Abstract
An alternative design for an amperometric biosensor for lactate is described in which the analyte flows through a rectangular duct without the aid of external pumping or a flow system. With this deviation from the traditional "stacked layers" of the biosensor geometry, the biorecognition matrix is located upstream of the detector electrode rather than directly over the electrode; this general design is reminiscent of the "dipstick" technology of the home test kits, but in the latter case the "flow" is a function of the nature of the wicking material rather than the dimensions of the channel. A droplet of analyte, l-lactate, is placed in the inlet well of the channel which flows through the cell and reacts with an immobilized lactate oxidase surface matrix zone, and the coproduct, hydrogen peroxide, is swept downstream to the monitoring electrode, where it is detected. The flow through the channel biosensor is shown to be laminar, and the biosensor response varies with the rate of flow of the analyte, which could be controlled by the outlet porosity; the sensitivity was observed to increase, but the dynamic range decreased with an increase in flow rate until a threshold loading, where the biosensor response became independent of the enzyme loading.Keywords
This publication has 19 references indexed in Scilit:
- Physical study of film-forming acrylate emulsion polymers for biosensor applicationsAnalytica Chimica Acta, 1997
- An enzyme electrode with response independent of the thickness of the enzyme layerSensors and Actuators B: Chemical, 1996
- Membrane properties of acrylate bulk polymers for biosensor applicationsBiosensors and Bioelectronics, 1996
- Electrochemical enzyme sensor for formaldehyde operating in the gas phaseBiosensors and Bioelectronics, 1996
- Challenges in integrating biosensors and FIA for on-line monitoring and controlTrends in Biotechnology, 1996
- Miniaturized thin film glutamate and glutamine biosensorsBiosensors and Bioelectronics, 1995
- Micromachined electrochemical flow cell for biosensingElectroanalysis, 1994
- Immobilisation of photosynthetic cells based on film-forming emulsion polymersAnalytica Chimica Acta, 1994
- An electrochemical capillary fill device for the analysis of glucose incorporating glucose oxidase and ruthenium (III) hexamine as mediatorElectroanalysis, 1992
- Model of a two-substrate enzyme electrode for glucoseAnalytical Chemistry, 1984