Control of parasitic aberrations in multipole optics
- 5 March 2009
- journal article
- Published by Oxford University Press (OUP) in Journal of Electron Microscopy
- Vol. 58 (3) , 123-130
Abstract
A method is described to find the optimal fourth order setup of a quadrupole–octupole third-order aberration corrector. Given accurate measurements of aberrations to fifth order, stimulus/response experiments can be used to synthesize pure controls for each measured aberration up to fourth order, including those which are caused by parasitic effects – symmetry violations, misalignments, construction mistakes, post-construction drift or other problems.Keywords
This publication has 10 references indexed in Scilit:
- Chapter 4 First Results Using the Nion Third-Order Scanning Transmission Electron Microscope Corrector2008
- Motion of Gold Atoms on Carbon in the Aberration-Corrected STEMMicroscopy and Microanalysis, 2007
- Characterizing probe performance in the aberration corrected STEMUltramicroscopy, 2006
- Prospects for aberration-free electron microscopyUltramicroscopy, 2005
- Aberration correction results in the IBM STEM instrumentUltramicroscopy, 2003
- Progress in aberration-corrected scanning transmission electron microscopyMicroscopy, 2001
- Electron microscopy image enhancedNature, 1998
- Towards 0.1 nm resolution with the first spherically corrected transmission electron microscopeJournal of Electron Microscopy, 1998
- Resolution enhancement by deconvolution using a field emission source in electron energy loss spectroscopyUltramicroscopy, 1992
- Forty Years of History of a Grating InterferometerApplied Optics, 1964