Assessment of EchoMRI-AH versus dual-energy X-ray absorptiometry to measure human body composition
Authors:
Open Access
- 11 January 2011
- journal article
- research article
- Published by Springer Nature in International Journal of Obesity
- Vol. 35 (9) , 1241-1246
Abstract
Background: The sensitivity to detect small changes in body composition (fat mass and fat-free mass) largely depends on the precision of the instrument. We compared EchoMRI-AH and dual-energy X-ray absorptiometry (DXA) (Hologic QDR-4500A) for estimating fat mass in 301 volunteers. Methods: Body composition was evaluated in 136 males and 165 females with a large range of body mass index (BMI) (19−49 kg m−2) and age (19−91 years old) using DXA and EchoMRI-AH. In a subsample of 13 lean (BMI=19–25 kg m−2) and 21 overweight/obese (BMI>25 kg m−2) individuals, within-subject precision was evaluated from repeated measurements taken within 1 h (n=3) and 1 week apart (mean of three measurements taken on each day). Results: Using Bland–Altman analysis, we compared the mean of the fat mass measurements versus the difference in fat mass measured by both instruments. We found that EchoMRI-AH quantified larger amount of fat versus DXA in non-obese (BMI−2 (1.1 kg, 95% confidence interval (CI95):−3.7 to 6.0)) and obese (BMI⩾30 kg m−2 (4.2 kg, CI95:−1.4 to 9.8)) participants. Within-subject precision (coefficient of variation, %) in fat mass measured within 1 h was remarkably better when measured by EchoMRI-AH than DXA (PP=0.15). Conclusions: EchoMRI-AH yielded greater fat mass values when compared with DXA (Hologic QDR-4500A), particularly in fatter subjects. EchoMRI-AH and DXA showed similar 1-week apart precision when fat mass was measured both in lean and overweight/obese individuals.Keywords
This publication has 12 references indexed in Scilit:
- When Prevention Fails: Obesity Treatment StrategiesThe American Journal of Medicine, 2009
- Validation of a Quantitative Magnetic Resonance Method for Measuring Human Body CompositionObesity, 2008
- Opportunities and challenges for the development of pharmacological therapies for obesity treatmentObesity Reviews, 2007
- QDR 4500A dual-energy X-ray absorptiometer underestimates fat mass in comparison with criterion methods in adultsThe American Journal of Clinical Nutrition, 2005
- Quantitative magnetic resonance (QMR) method for bone and whole-body-composition analysisAnalytical and Bioanalytical Chemistry, 2003
- Obesity as a medical problemNature, 2000
- Within- and Between-Subject Variation in Commonly Measured Anthropometric and Biochemical VariablesClinical Chemistry, 1999
- Measurement of soft tissue composition by dual energy X-ray absorptiometryBritish Journal of Nutrition, 1997
- Dual-Energy X-Ray Absorptiometry Measurements of Body Composition: Effects of Depth and Tissue Thickness, Including Comparisons with Direct AnalysisClinical Science, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986