GlucoMen Day Continuous Glucose Monitoring System: A Screening for Enzymatic and Electrochemical Interferents
Open Access
- 1 September 2012
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 6 (5) , 1172-1181
Abstract
Background: While most of the common drugs with the potential to interfere with continuous glucose monitoring (CGM) systems are accessible over the counter and can be assumed by CGM patients without medical supervision, many other chemicals are frequently used to treat critically ill patients. Continuous glucose monitoring reading accuracy may also be compromised in patients characterized by abnormally high concentrations of physiological interferents. In this article, 22 species selected from endogenous and exogenous chemicals were screened as possible interferents of GlucoMen®Day (GMD), the new microdialysis-based CGM system from A. Menarini Diagnostics. Method: Interference testing was performed according to the EP7-A2 guideline (Clinical and Laboratory Standards Institute 2005). Interference was evaluated at two levels of glucose, with each interferent additionally tested at two concentrations. Furthermore, two configurations of the GMD disposable sensor kit—one designed for subcutaneous application, the other for direct intravascular CGM—were challenged with interferent-spiked serum and blood samples, respectively. Results: With the exception of dopamine (however, at very high, nonphysiological concentrations), no interference was observed for all the tested substances. Interestingly, none of the common electrochemical interferents (including ascorbic acid, acetaminophen, and salicylic acid, which represent the major specificity issue for the competing CGM systems) significantly affected the system's output. Conclusions: These results provide clear insights into the advantages offered by the use of a microdialysis-based CGM system that additionally relies on the detection of hydrogen peroxide at low operating potential. GlucoMen Day may become the CGM system of choice for those patients who require either regular administration of drugs or their glycemia to be tightly controlled in the intensive care unit or similar environments.Keywords
This publication has 15 references indexed in Scilit:
- Evaluating the Clinical Accuracy of GlucoMen®Day: A Novel Microdialysis-Based Continuous Glucose MonitorJournal of Diabetes Science and Technology, 2010
- The Chemistry of Commercial Continuous Glucose MonitorsDiabetes Technology & Therapeutics, 2009
- Pre- and Postoperative Accuracy and Safety of a Real-Time Continuous Glucose Monitoring System in Cardiac Surgical Patients: A Randomized Pilot StudyDiabetes Technology & Therapeutics, 2009
- Ex Vivo Continuous Glucose Monitoring With Microdialysis Technique: The Example of GlucoDayIEEE Sensors Journal, 2008
- Electrochemical Glucose BiosensorsChemical Reviews, 2007
- Toward continuous glucose monitoring with planar modified biosensors and microdialysisBiosensors and Bioelectronics, 2007
- Intensive Insulin Therapy in the Intensive Care UnitDiabetes Care, 2006
- Novel planar glucose biosensors for continuous monitoring useBiosensors and Bioelectronics, 2005
- A Continuous Glucose Sensor Based on Wired Enzyme ™ Technology - Results from a 3-Day Trial in Patients with Type 1 DiabetesDiabetes Technology & Therapeutics, 2003
- Intensive Insulin Therapy in Critically Ill PatientsNew England Journal of Medicine, 2001