Repeatability of exhaled breath fingerprint collected by a modern sampling system in asthmatic and healthy children
- 1 May 2019
- journal article
- research article
- Published by IOP Publishing in Journal of Breath Research
- Vol. 13 (3) , 036007
Abstract
E-noses provide potential non-invasive metabolic biomarkers for diagnosing and monitoring pulmonary diseases. The primary aim of the present study was to assess the within-day and between-day repeatability of a modern breath sampling system (Pneumopipe ® plus an array of e-nose sensors) in asthmatic and healthy children. The secondary aim was to compare the repeatability of the breath sampling system, spirometry and exhaled nitric oxide (eNO). Fifteen children (age 6–11 years) with asthma and thirty healthy children matched by age and gender (1:2 allocation) were recruited; of them, three healthy children did not complete the study. All measurements were collected twice during the baseline visit, 30 min apart, and once during the final visit, after 7 d. Repeatability was assessed through the intra-cluster correlation coefficient (ICC), and a significance test was performed to detect an at least ‘fair’ repeatability (ICC > 0.2). In asthmatic children, the within-day (0–30 min) ICCs for e-nose sensors (8 sensors × 4 desorption temperatures) ranged from 0.24 to 0.84 (median 0.57, IQR 0.47-0.71), while the between-day (0–7 d) ICCs ranged from 0.25 to 0.83 (median 0.66, IQR 0.55-0.72). In healthy children, the within-day ICCs for e-nose sensors ranged from 0.29 to 0.85 (median 0.58, IQR 0.49-0.63), while the between-day ICCs ranged from 0.33 to 0.82 (median 0.55, IQR 0.49-0.63). In both groups, most of the within-day and between-day ICCs for e-nose sensors were statistically significant. Moreover, the within-day and between-day ICCs for all spirometry parameters and eNO were significant and similar to those of the most reliable sensors. The modern breath sampling system showed more than acceptable within-day and between-day repeatability, in both asthmatic and healthy children. The present study was registered on the central registration system ClinicalTrials.gov (ID: NCT03025061).Keywords
This publication has 31 references indexed in Scilit:
- Design and Test of a Biosensor-Based Multisensorial System: A Proof of Concept StudySensors, 2013
- Within-day and between-day repeatability of measurements with an electronic nose in patients with COPDJournal of Breath Research, 2013
- The Electronic Nose in Respiratory MedicineRespiration, 2012
- Proportional Venn diagram and determinants of allergic respiratory diseases in Italian adolescentsPediatric Allergy and Immunology, 2010
- Strictly Standardized Mean Difference, Standardized Mean Difference and Classicalt-test for the Comparison of Two GroupsStatistics in Biopharmaceutical Research, 2010
- Diagnostic Performance of an Electronic Nose, Fractional Exhaled Nitric Oxide, and Lung Function Testing in AsthmaChest, 2010
- Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variablesUltrasound in Obstetrics & Gynecology, 2008
- Development and cross-sectional validation of the Childhood Asthma Control TestJournal of Allergy and Clinical Immunology, 2007
- Standardisation of spirometryEuropean Respiratory Journal, 2005
- Clinical and Biological Heterogeneity in Children with Moderate AsthmaAmerican Journal of Respiratory and Critical Care Medicine, 2003