A New Look at Calibration and Use of Eppley Precision Infrared Radiometers. Part II: Calibration and Use of the Woods Hole Oceanographic Institution Improved Meteorology Precision Infrared Radiometer*
For some years, investigators have made measurements of downwelling longwave irradiance with the Eppley Precision Infrared Radiometer (PIR), recording the values of thermopile voltage and body and dome thermistor resistances and combining them in data processing. Part I of this paper reviews previous work on the processing equation and presents an improved equation. It establishes that the standard single-output Eppley has an inherent uncertainty of 5%. By measuring the three possible outputs separately and comparing them in the improved equation, the inherent accuracy can be improved to 1.5%. Part II presents a method of calibrating the Eppley PIR for the three-output equation using an easily constructed blackbody cavity in a temperature bath capable of a 0°–50°C temperature range. Calibration of PIR thermistors is recommended since occasionally one is found out of specifications. An outdoor comparison of 15 PIRs calibrated with the technique was carried out in groups of four, with one PIR used ... Abstract For some years, investigators have made measurements of downwelling longwave irradiance with the Eppley Precision Infrared Radiometer (PIR), recording the values of thermopile voltage and body and dome thermistor resistances and combining them in data processing. Part I of this paper reviews previous work on the processing equation and presents an improved equation. It establishes that the standard single-output Eppley has an inherent uncertainty of 5%. By measuring the three possible outputs separately and comparing them in the improved equation, the inherent accuracy can be improved to 1.5%. Part II presents a method of calibrating the Eppley PIR for the three-output equation using an easily constructed blackbody cavity in a temperature bath capable of a 0°–50°C temperature range. Calibration of PIR thermistors is recommended since occasionally one is found out of specifications. An outdoor comparison of 15 PIRs calibrated with the technique was carried out in groups of four, with one PIR used ...