Feasibility, safety and accuracy of a CT-guided robotic assistance for percutaneous needle placement in a swine liver model
Authors:
Open Access
- 4 March 2021
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 11 (1) , 1-11
Abstract
Evaluate the feasibility, safety and accuracy of a CT-guided robotic assistance for percutaneous needle placement in the liver. Sixty-six fiducials were surgically inserted into the liver of ten swine and used as targets for needle insertions. All CT-scan acquisitions and robotically-assisted needle insertions were coordinated with breath motion using respiratory monitoring. Skin entry and target points were defined on planning CT-scan. Then, robotically-assisted insertions of 17G needles were performed either by experienced interventional radiologists or by a novice. Post-needle insertion CT-scans were acquired to assess accuracy (3D deviation, ie. distance from needle tip to predefined target) and safety. All needle insertions (43/43; median trajectory length = 83 mm (interquartile range [IQR] 72–105 mm) could be performed in one (n = 36) or two (n = 7) attempts (100% feasibility). Blinded evaluation showed an accuracy of 3.5 ± 1.3 mm. Accuracy did not differ between novice and experienced operators (3.7 ± 1.3 versus 3.4 ± 1.2 mm, P = 0.44). Neither trajectory angulation nor trajectory length significantly impacted accuracy. No complications were encountered. Needle insertion using the robotic device was shown feasible, safe and accurate in a swine liver model. Accuracy was influenced neither by the trajectory length nor by trajectory angulations nor by operator’s experience. A prospective human clinical trial is recruiting.Keywords
Funding Information
- Quantum Surgical
This publication has 24 references indexed in Scilit:
- Multimodal Percutaneous Thermal Ablation of Small Hepatocellular Carcinoma: Predictive Factors of Recurrence and Survival in Western PatientsCancers, 2020
- Percutaneous thermal ablation of hepatocellular carcinomas located in the hepatic dome using artificial carbon dioxide pneumothorax: retrospective evaluation of safety and efficacyInternational Journal of Hyperthermia, 2018
- Percutaneous treatment of hepatocellular carcinoma: State of the art and innovationsJournal of Hepatology, 2018
- Robotic-assisted thermal ablation of liver tumoursEuropean Radiology, 2014
- Real-Time US-CT/MRI Image Fusion for Guidance of Thermal Ablation of Liver Tumors Undetectable with US: Results in 295 CasesCardioVascular and Interventional Radiology, 2014
- Ten-year outcomes of percutaneous radiofrequency ablation as first-line therapy of early hepatocellular carcinoma: Analysis of prognostic factorsJournal of Hepatology, 2013
- Percutaneous radiofrequency ablation of liver cancer in the hepatic dome using the intrapleural fluid infusion techniqueBritish Journal of Surgery, 2008
- Thinking Outside the Abdominal Box: Safe Use of the Epipericardial Fat Pad Window for Percutaneous Radiofrequency Ablation of Hepatic Dome TumorsJournal of Vascular and Interventional Radiology, 2008
- Radiofrequency ablation for hepatocellular carcinoma in so-called high-risk locationsHepatology, 2006
- The Inaccessible or Undrainable Abscess: How to Drain ItRadioGraphics, 2004