Clinical results with Jarvik 2000 axial flow left ventricular assist device: Osaka University Experience
- 22 July 2014
- journal article
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 17 (4) , 308-314
Abstract
The aim of this study was to evaluate our clinical experience with the Jarvik 2000 axial flow pump (Jarvik Heart, Inc, New York, NY, USA), a miniature axial flow left ventricular assist device (LVAD). The clinical results of eight patients, who underwent LVAD implantation with the Jarvik 2000 (median age 55.0 years; six men) between 2005 and 2010, including two who participated in a multicenter clinical trial in Japan, were reviewed. Two patients underwent LVAD implantation as destination therapy. Four patients underwent Jarvik 2000 implantation via median sternotomy, while the other four underwent implantation via left thoracotomy. There were no major complications during surgery. Four patients were supported for more than 2 years. The longest support duration was 1,618 days. Six patients successfully bridged to heart transplantation after a median 725 days of support. One patient on destination therapy died of a cerebral infarction. The other patient on destination therapy had had the LVAD for 1,618 days. The overall survival rates at 1, 2, and 3 years were 100, 86, and 86 %, respectively. The median postoperative serum lactate dehydrogenase level was 860.5 U/L at 1 month, 735 U/L at 6 months, and 692 U/L at 1 year. There were no fatal device-related infections. We found that the Jarvik 2000 with pin bearing could support patients with end-stage heart failure with acceptable mortality and morbidity rates. Further evaluations of the prevalence of thromboembolic and hemolytic events in patients with the new conical-bearing Jarvik 2000 are required.Keywords
This publication has 17 references indexed in Scilit:
- Outcomes of pediatric patients supported by the HeartMate II left ventricular assist device in the United StatesThe Journal of Heart and Lung Transplantation, 2013
- Clinical Differences Between Continuous Flow Ventricular Assist Devices: A Comparison Between HeartMate II and HeartWare HVADJournal of Cardiac Surgery, 2013
- DuraHeartTM Magnetically Levitated Left Ventricular Assist DeviceCirculation Journal, 2013
- Lower body surface area is highly related to mortality due to stroke or systemic bleeding in patients receiving an axial flow blood pump as a left ventricular assist deviceEuropean Journal of Cardio-Thoracic Surgery, 2012
- Pre-operative and post-operative risk factors associated with neurologic complications in patients with advanced heart failure supported by a left ventricular assist deviceThe Journal of Heart and Lung Transplantation, 2012
- University of Maryland Surgical Experience With the Jarvik 2000 Axial Flow Ventricular Assist DeviceThe Annals of Thoracic Surgery, 2012
- The first clinical case in Japan of destination therapy using the Jarvik 2000 left ventricular assist deviceJournal of Artificial Organs, 2010
- Evaluation of platelet activation in patients supported by the Jarvik 2000∗ high–rotational speed impeller ventricular assist deviceThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Mechanical Reliability of the Jarvik 2000 HeartThe Annals of Thoracic Surgery, 2006
- Initial Clinical Experience With the Jarvik 2000 Implantable Axial-Flow Left Ventricular Assist SystemCirculation, 2002