Clinical Application of a Digital Computer Software System to Electrodiagnosis

Authors:
  • YMYozo Miyake
    Yozo Miyake
  • ASAlfred Solish
    Alfred Solish
  • AHAkira Hara
    Akira Hara
  • THTatsuo Hirose
    Tatsuo Hirose

Abstract
A digital computer software system was designed for improved recording and analyzing electrical responses from the human visual system in a clinical setting. Three sample procedures that were developed demonstrate the capability and flexibility of the software system: (1) selection of artifact-reduced responses for averaging. A maximum voltage was set in the computer program which, if exceeded, caused the trace containing the aberrant voltage to be excluded from the average. The maximum tolerance test can be applied to any portion and any duration of the tracing. This feature is particularly valuable when a large artifact is expected in each stimulus delivery, as seen in the electrically evoked response. (2) Measurement of amplitude and peak time using two cursors displayed on a cathode ray tube screen. (3) Isolation of the oscillatory potential from the total ERG without the distortion introduced by the b-wave. These techniques, using a computer software system, help the ophthalmologist to analyze the electrical response from the human visual system quickly, easily, and with a high degree of precision.