Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery
Authors:
- 23 March 2019
- journal article
- research article
- Published by Springer Nature in Journal of Anesthesia
- Vol. 33 (3) , 364-371
Abstract
Purpose The ClearSight™ device monitors continuous pressure and cardiac output via pulse contour analysis. ClearSight™, however, may not be reliable in patients with reduced peripheral perfusion caused by high peripheral resistance. This study aimed to elucidate the accuracy and trending ability of ClearSight™ in patients undergoing abdominal aortic aneurysm (AAA) surgery by comparing the ClearSightTM-derived cardiac index (CICS) with that measured using three-dimensional echocardiography (CI3D). Methods The study included 20 patients who underwent elective AAA surgery. CICS and CI3D were measured simultaneously at eight time points during the surgery. Trending ability was investigated after aortic clamping and unclamping. We used CI3D as the reference method. Results Bland–Altman analysis showed a wide limit of agreement between CICS and CI3D (percentage error 41.3%). Subgroup analysis showed a lower percentage error (33.2%) in patients with CI ≥ 2.5 L/min/m2. The cubic splines related to the CI3D and CI discrepancy were negatively sloped, indicating that CI3D had significant influence on the CI discrepancy (p < 0.001). Four-quadrant plot analysis showed that the tracking ability of ClearSight™ after aortic clamping and declamping were clinically unacceptable (81.3% and 78.6%, respectively). Also, the polar plot analysis showed that the concordance rate of ClearSight™ after aortic clamping and declamping were clinically unacceptable (58.3% and 66.7%, respectively). Conclusions ClearSight™ was not sufficiently accurate in patients undergoing AAA surgery. The tracking ability of ClearSight™ after aortic clamping was below the acceptable limit.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP17K11100)
This publication has 28 references indexed in Scilit:
- Assessment of changes in cardiac index and fluid responsiveness: a comparison of Nexfin and transpulmonary thermodilutionActa Anaesthesiologica Scandinavica, 2013
- Comparative Assessment of 2-Dimensional Echocardiography vs Cardiac Magnetic Resonance Imaging in Measuring Left Ventricular Mass in Patients With and Without End-Stage Renal DiseaseCanadian Journal of Cardiology, 2013
- Investigation of the freely available easy-to-use software ‘EZR’ for medical statisticsBone Marrow Transplantation, 2012
- The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challengeCritical Care, 2012
- Clinical Comparison of an Echocardiograph-Derived Versus Pulse Counter–Derived Cardiac Output Measurement in Abdominal Aortic Aneurysm SurgeryJournal of Cardiothoracic and Vascular Anesthesia, 2012
- A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac OutputAnesthesia & Analgesia, 2010
- Real-Time 3-Dimensional EchocardiographyCirculation, 2009
- Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiographyJACC, 2004
- Recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and doppler echocardiographyJournal of the American Society of Echocardiography, 2003
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986