Measurement of meniscofemoral contact pressure after repair of bucket-handle tears with biodegradable implants
Authors:
- 7 September 2004
- journal article
- research article
- Published by Springer Nature in Archives of orthopaedic and trauma surgery
- Vol. 125 (4) , 254-260
Abstract
Introduction Biodegradable implants are frequently used for meniscus repair. Articular cartilage damage has been reported recently after meniscus repair with biodegradable implants. The aim of the study was to investigate the meniscofemoral contact pressure at the posterior horn of the medial and lateral meniscus after repair of bucket-handle lacerations. Materials and methods Specimens were mounted in a materials testing machine (Bionix 858, MTS) which was equipped with a load cell. The quadriceps tendon was attached to a hydraulic cylinder, and knee motion was controlled via tension of the quadriceps tendon. A piezo-resistive system (Tekscan, Boston, MA, USA) measured the meniscofemoral contact pressure. Five different types of biodegradable implants (Arrow, Dart, Fastener, Stinger and Meniscal Screw) and horizontal suture (no. 2 Ethibond) were tested. The knee was extended from 90° of flexion to 0° under a constant load of 350 N due to adjustment of the tension force of the quadriceps tendon. The femorotibial pressure and contact area were recorded at 0°, 30°, 60° and 90° of flexion. Results The meniscofemoral pressure did not increase after meniscus repair with biodegradable implants or sutures. The meniscofemoral peak pressure at the posterior horn was 1.46±1.54 MPa in the medial compartment and 1.08±1.17 MPa in the lateral compartment at full knee extension. The meniscofemoral pressure increased significantly in both compartments with knee flexion from 0° to 90°. Conclusion Biodegradable implants for meniscus repair do not affect the meniscofemoral pressure. However, there remains a risk of damage to the cartilage when barbed implants are used. If the implant is not entirely advanced into the meniscus, the sharp head or some of the barbs at the column of the implant may come into direct contact with the articular cartilage of the femoral condyle or tibial plateau. The authors presume that incorrect positioning of the implant seems to be the major reason for cartilage damage.Keywords
This publication has 35 references indexed in Scilit:
- Biomechanical Properties Under Cyclic Loading of Seven Meniscus Repair TechniquesClinical Orthopaedics and Related Research, 2002
- An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji filmJournal of Biomechanics, 1999
- Struktur und Blutgefäßversorgung der KniegelenkmeniskenZeitschrift für Orthopädie und ihre Grenzgebiete, 1999
- Long Term Results After Arthroscopic Meniscectomy: The Role of Preexisting Cartilage Fibrillation in a 13 Year Follow-Up of 60 PatientsInternational Journal of Sports Medicine, 1996
- Tensile stress-strain characteristics of the human meniscal materialJournal of Biomechanics, 1995
- Radial tie fibers influence the tensile properties of the bovine medial meniscusJournal of Orthopaedic Research, 1994
- Partial meniscectomy and anterior cruciate ligament rupture in soccer playersThe American Journal of Sports Medicine, 1993
- Materialeigenschaften von Meniskusgewebe sowie der Einfluß der Menisken auf das Kompressionsverhalten des KniegelenkesZeitschrift für Orthopädie und ihre Grenzgebiete, 1993
- A transversely isotropic biphasic finite element model of the meniscusJournal of Biomechanics, 1992
- A Pressure Distribution Transducer for In-Vitro Static Measurements in Synovial JointsJournal of Biomechanical Engineering, 1983