Comparison Between Two On-Line Reversed Line Position Hemodialysis Vascular Access Flow Measurement Techniques: Saline Dilution and Thermodilution
Authors:
- Stig EssersSEStig Essers
- Ger van MeijelGMGer van Meijel
- Jeroen P. KoomanJKJeroen P. Kooman
- Jan TordoirJTJan Tordoir
- Karel M.L. LeunissenKLKarel M.L. Leunissen
- Frank M. van der SandeFSFrank M. van der Sande
- 1 July 2006
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 52 (4) , 410-415
Abstract
Periodical access flow measurements can predict the development and presence of vascular access flow-limiting stenosis and subsequent thrombosis. Access flow measurement has become a standard in vascular access care. Different techniques to measure access flow are available. The aim of this study was to compare an integrated access flow measurement device, based on thermodilution (Blood Temperature Monitor, BTM, Fresenius Medical Care, Bad Homburg, Germany), with the gold standard, the HD01 (Transonic Systems Inc., Ithaca, NY), whose technique is based on saline dilution. In 40 patients with end-stage renal disease, 40 vascular accesses were studied to determine the correlation between access flow measurements by both techniques. Reproducibility of access flow measurements by both techniques was assessed in 20 patients on a weekly interval.A total of 40 measurement series were performed. Average access flow measured with the saline technique and the thermodilution technique was 1053 (+/-495) ml/min and 1034 (+/-527) ml/min, respectively (p = ns) (n = 40). Correlation between access flow measurements by both techniques expressed in R was 0.79 (r = 0.89). Reproducibility of saline and thermodilution subsequent measurements with a weekly interval, expressed in relative difference (Delta xrel) was 13 (+/-11)% and 24 (+/-14)%, respectively (p < 0.01) (n = 20).BTM access flow measurements correlated well with the HD01 access flow measurements. However, the better reproducibility of HD01 and shorter measurement time compared with BTM access flow measurements should be considered when implementing access flow measurement to prevent vascular access failure.Keywords
This publication has 14 references indexed in Scilit:
- Access Monitoring Does Not Really Improve OutcomesBlood Purification, 2004
- Vascular Access Monitoring Improves OutcomesBlood Purification, 2004
- Comparison of Different Techniques of Hemodialysis Vascular Access Flow EvaluationThe International Journal of Artificial Organs, 2003
- Vascular access blood flow monitoring reduces access morbidity and costsKidney International, 2001
- Hemodialysis arteriovenous access: Detection of stenosis and response to treatment by vascular access blood flowKidney International, 2001
- III. NKF-K/DOQI Clinical Practice Guidelines for Vascular Access: Update 2000American Journal of Kidney Diseases, 2001
- Intervention Based on Monthly Monitoring Decreases Hemodialysis Access ThrombosisASAIO Journal, 1999
- Validation of haemodialysis recirculation and access blood flow measured by thermodilutionNephrology Dialysis Transplantation, 1999
- Measurement of Access Flow During Hemodialysis Using the Constant Infusion ApproachASAIO Journal, 1998
- Theory and validation of access flow measurement by dilution technique during hemodialysisKidney International, 1995