Evaluation and Implementation of the Gel Test for Indirect Antiglobulin Testing in a Community Hospital Laboratory
Authors:
Open Access
- 1 August 1999
- journal article
- Published by Archives of Pathology and Laboratory Medicine in Archives of Pathology & Laboratory Medicine
- Vol. 123 (8) , 693-697
Abstract
Background. —The gel test, developed by Lapierre in 1984, was designed to standardize antiglobulin testing while improving sensitivity and specificity of the method. Principle. —Anti–human serum immunoglobulin G (IgG) mixed with Sephadex G100 (gel phase) in a microtube traps red cell–IgG agglutination complexes during migration through the gel in a centrifugation step. Agglutination complexes are visibly detectable at various levels in the microtube as an inverse function of antibody coated on red cells. Unsensitized red cells form a cell pellet at the base of the microtube. Objective. —To determine if indirect anti–human globulin testing could be standardized and simplified by replacing the tube test with the gel test without compromising quality or increasing costs. Setting. —A medium-sized community hospital. Results. —In a blinded retrospective study, we used patient sera (n = 40), which included 10 positive specimens containing 18 known antibodies. Sixteen antibodies were detected and identified with the tube method (1 anti-D and 1 anti-C not detected). By the gel test, 18 antibodies were detected and identified. All negative samples showed 100% concordance. Favorable results were obtained in a nonblinded prospective correlation study (n = 121). Our technologists found the gel test easier to read and more reproducible and reliable than the tube method; they also found increased sensitivity for detecting weakly reacting antibodies. We successfully introduced the gel test into our laboratory as the standard method for indirect antiglobulin testing. Following implementation, improved personnel management was achieved. Conclusions. —The gel test is a reliable and advantageous method and is appropriate for routine use for detection and identification of alloantibodies in a community hospital transfusion service laboratory.Keywords
This publication has 15 references indexed in Scilit:
- Unsatisfactory Detection of an in vivo Haemolytic Anti‐Vel by the Gel TestVox Sanguinis, 1998
- Historical Note: Past, Present and Future of the Antiglobulin TestVox Sanguinis, 1998
- Evaluation of the ID‐gel test for antibody screening and identificationTransfusion Medicine, 1991
- The role of compatibility tests. (Report of a meeting sponsored by the Bureau of Biologics for the Blood Products Advisory Committee)Transfusion, 1982
- Pretransfusion Blood Group SerologyTransfusion, 1979
- The Risk of Abbreviating the Major Crossmatch in Urgent or Massive TransfusionTransfusion, 1978
- The type and screen: a safe alternative and supplement in selected surgical proceduresTransfusion, 1977
- The Maximum Surgical Blood Order Schedule and Surgical Blood Use in the United StatesTransfusion, 1976
- Routine Compatibility Testing Standards of the AABB as Applied to Compatibility TestsTransfusion, 1964
- DETECTION OF WEAK AND "INCOMPLETE" Rh AGGLUTININS: A NEW TESTThe Lancet, 1945