Determination of electron energy, spectral width, and beam divergence at the exit window for clinical megavoltage x‐ray beams
Authors:
- D. L. SawkeyDSD. L. Sawkey
- B. A. FaddegonBFB. A. Faddegon
- 3 February 2009
- journal article
- Published by Wiley in Medical Physics
- Vol. 36 (3) , 698-707
Abstract
Monte Carlo simulations of x‐ray beams typically take parameters of the electron beam in the accelerating waveguide to be free parameters. In this paper, a methodology is proposed and implemented to determine the energy, spectral width, and beam divergence of the electron source. All treatment head components were removed from the beam path, leaving only the exit window. With the x‐ray target and flattener out of the beam, uncertainties in physical characteristics and relative position of the target and flattening filter, and in spot size, did not contribute to uncertainty in the energy. Beam current was lowered to reduce recombination effects. The measured dose distributions were compared with Monte Carlo simulation of the electron beam through the treatment head to extract the electron source characteristics. For the nominal 6 and x‐ray beams, the energies were and , respectively, with the uncertainties resulting from uncertainties in the detector position in the measurement and in the stopping power in the simulations. Gaussian spectral distributions were used, with full widths at half maximum ranging from at to at required to match the fall‐off portion of the percent‐depth ionization curve. Profiles at the depth of maximum dose from simulations that used the manufacturer‐specified exit window geometry and no beam divergence were narrower than measured profiles. Two simulation configurations yielding the measured profile width were the manufacturer‐specified exit window thickness with electron source divergences of 3.3° at and 1.8° at and an exit window 40% thicker than the manufacturer's specification with no beam divergence. With the x‐ray target in place (and no flattener), comparison of measured to simulated profiles sets upper limits on the electron source divergences of 0.2° at and 0.1° at . A method of determining source characteristics without mechanical modification of the treatment head, and therefore feasible in clinics, is presented. The energies and spectral widths determined using this method agree with those determined with only the exit window in the beam path.Keywords
Funding Information
- National Institutes of Health (R01 CA104777-01A2)
This publication has 29 references indexed in Scilit:
- The effective point of measurement of ionization chambers and the build‐up anomaly in MV x‐ray beamsMedical Physics, 2008
- Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo‐based photon and electron external beam treatment planningMedical Physics, 2007
- On the discrepancies between Monte Carlo dose calculations and measurements for the Varian photon beamMedical Physics, 2007
- Determination of the initial beam parameters in Monte Carlo linac simulationMedical Physics, 2006
- Decoupling initial electron beam parameters for Monte Carlo photon beam modelling by removing beam-modifying filters from the beam pathPhysics in Medicine & Biology, 2005
- Monte Carlo modelling of external radiotherapy photon beamsPhysics in Medicine & Biology, 2003
- Sensitivity of megavoltage photon beam Monte Carlo simulations to electron beam and other parametersMedical Physics, 2002
- Monte Carlo modelling of electron beams from medical acceleratorsPhysics in Medicine & Biology, 1999
- Angular distribution of bremsstrahlung from 15‐MeV electrons incident on thick targets of Be, Al, and PbMedical Physics, 1991
- Forward‐directed bremsstrahlung of 10‐ to 30‐MeV electrons incident on thick targets of Al and PbMedical Physics, 1990