Optimal Force–Time Integral for Pulmonary Vein Isolation According to Anatomical Wall Thickness Under the Ablation Line
Open Access
- 9 March 2016
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of the American Heart Association
- Vol. 5 (3) , e003155
Abstract
Background: Low contact force and force–time integral ( FTI ) during catheter ablation are associated with ineffective lesion formation, whereas excessively high contact force and FTI may increase the risk of complications. We sought to evaluate the optimal FTI for pulmonary vein ( PV ) isolation based on atrial wall thickness under the ablation line. Methods and Results: Contact force parameters and FTI during anatomical ipsilateral PV isolation for atrial fibrillation and atrial wall thickness were assessed retrospectively in 59 consecutive patients for their first PV isolation procedure. The PV antrum was divided into 8 segments, and the wall thickness of each segment under the ablation line was determined using multidetector computed tomography. The FTI for each ablation point was divided by the wall thickness of the PV antrum segment where each point was located to obtain FTI /wall thickness. In total, 5335 radiofrequency applications were delivered, and 85 gaps in PV isolation ablation lines and 15 dormant conductions induced by adenosine were detected. The gaps or dormant conductions were significantly associated with low contact force, radiofrequency duration, FTI , and FTI /wall thickness. Among them, FTI /wall thickness had the best prediction value for gaps or dormant conductions by receiver operating characteristic curve analysis. FTI /wall thickness of <76.4 gram‐seconds per millimeter (gs/mm) predicted gaps or dormant conductions with sensitivity (88.0%) and specificity (83.6%), and FTI /wall thickness of <101.1 gs/mm was highly predictive (sensitivity 97.0%; specificity 69.6%). Conclusions: FTI /wall thickness is a strong predictor of gap and dormant conduction formation in PV isolation. An FTI /wall thickness ≈100 gs/mm could be a suitable target for effective ablation.Keywords
This publication has 18 references indexed in Scilit:
- EFFICAS II: optimization of catheter contact force improves outcome of pulmonary vein isolation for paroxysmal atrial fibrillationEP Europace, 2015
- Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesionsEP Europace, 2014
- Electrical Reconnection After Pulmonary Vein Isolation Is Contingent on Contact Force During Initial TreatmentCirculation: Arrhythmia and Electrophysiology, 2013
- Left atrial thickness under the catheter ablation lines in patients with paroxysmal atrial fibrillation: insights from 64-slice multidetector computed tomographyHeart and Vessels, 2012
- Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force‐Sensing CatheterJournal of Cardiovascular Electrophysiology, 2010
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- A Systematical Analysis of In Vivo Contact Forces on Virtual Catheter Tip/Tissue Surface Contact during Cardiac Mapping and InterventionJournal of Cardiovascular Electrophysiology, 2008
- The architecture of the left lateral atrial wall: a particular anatomic region with implications for ablation of atrial fibrillationEuropean Heart Journal, 2008
- External and Endoluminal Analysis of Left Atrial Anatomy and the Pulmonary Veins in Three‐Dimensional Reconstructions of Magnetic Resonance Angiography: The Full Insight from InsideJournal of Cardiovascular Electrophysiology, 2006
- Atrio-Esophageal Fistula as a Complication of Percutaneous Transcatheter Ablation of Atrial FibrillationCirculation, 2004