Course of von Willebrand factor in recipients of axial-flow and centrifugal blood pumps in up to 12 months of follow-up

Authors:
  • NDN Dranishnikov
    N Dranishnikov
  • ASA Stepanenko
    A Stepanenko
  • AFA Frumkin
    A Frumkin
  • JVJ Vierecke
    J Vierecke
  • EPEv Potapov
    Ev Potapov
  • TKT Krabatsch
    T Krabatsch
  • RHR Hetzer
    R Hetzer

Abstract
Aims: Von Willebrand factor (vWF) is one of the key factors in the coagulation system. The impact of different types of continuous left ventricular assist devices (LVADs) on vWF in mid-term follow-up is unknown. Methods: Between January 2010 and August 2011 blood samples were prospectively collected from 62 consecutive LVAD recipients (centrifugal flow group, HeartWare HVAD, n=44; axial flow group, HeartMate II, n=18). We analyzed levels of vWF antigen (VWF:AG), collagen-binding ability (vWF:CBA), CBA/AG ratio and presence of high-molecular weight multimers before and 7 days, 1, 3, 6 and 12 months after surgery. Results: There were no significant differences between groups in preoperative demographic parameters. The significant decrease (p0.05) observed between the two groups in the parameters measured preoperatively or during the whole period of follow-up. Conclusion: Implantation of a rotary blood pump as a left ventricular assist device leads to significant changes in von Willebrand factor profile, already after the first week of support. The type of blood pump (centrifugal or axial-flow) does not seem to play a significant role in these changes.