Abstract
In the given present study, the effect of pre-irradiation heat treatment at 500 and 600 °C on the glow peaks of synthetic quartz was examined as a function of annealing time to obtain an optimum annealing procedure. It was observed that the annealing time is not a strongly sensitive parameter to change the intensities of glow peaks. On the other hand, the intensities of glow peaks between room temperature (RT) and 200 °C were continuously increased during successive readings after heat treatments. Moreover, the intensities of glow peaks above 250 °C have good stabilities. The obtained repeatability of a glow peak at ∼320 °C over 10 cycles is within 5% after the application of annealing at 600 °C for 1 h. The general thermoluminescent dosimetric characteristics of synthetic quartz, such as the dose–response, signal fading as a function of storage time, and reusability were also tested using the annealing condition at 600 °C for 1 h. It was observed that dose-response behaviours of all glow peaks are similar to each other. They first follow linear part and then saturated at different dose levels. Peak 1 completely disappeared after 1 month storage in the dark room at RT. On the other hand, the intensity of peaks 2+3 was approximately reduced to 15% of its original value whereas the other peaks (P4–P5) were not sufficiently affected during this period.