Hot-wire anemometer for spirography
- 1 January 1998
- journal article
- respiratory measurement
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 36 (1) , 17-21
Abstract
The use of a constant temperature hot-wire anemometer flow sensor for spirography is reported. The construction, operating principles and calibration procedure of the apparatus are described, and temperature compensation method is discussed. Frequency response is studied. It is shown that this hot-wire flow transducer satisfies common demands with respect to accuracy, response time and temperature variations.Keywords
This publication has 11 references indexed in Scilit: