After a brief discussion of the past use of and clinical need for high frequency hearing tests, a new technique is described that uses a special electric transduction mode: a 60-kHz carrier frequency is modulated by the desired audiofrequency. The combined signal is applied to the skin over the mastoids by Mylar-covered electrodes. On perception, the audiofrequency part of the input shows all the properties of a bone-conduction signal. The carrier frequency is dispersed in the tissues. Production and calibration of electrical high frequency signals, up to 20 kHz, is straightforward and easier than that of conventional air-conduction and (vibromechanical) bone-conduction signals. The efficacy of the method is illustrated by means of preliminary hearing tests conducted on several subjects.Keywords
This publication has 15 references indexed in Scilit:
- Ear-canal resonances and the assessment of hearing thresholds at high frequenciesThe Journal of the Acoustical Society of America, 1982
- Behaviour of High-Frequency Hearing in NoiseInternational Journal of Audiology, 1982
- Bone Conduction1976
- The External Ear1974
- Effects Of Noise Upon The Upper Frequency Limit Of HearingActa Oto-Laryngologica, 1973
- Hearing Rehabilitation By Means Of Transdermal Electrotherapy In Human Hearing Loss Of Sensorineural Origin: Preliminary ReportActa Oto-Laryngologica, 1969
- Bone-Conduction Thresholds for Sonic and Ultrasonic FrequenciesThe Journal of the Acoustical Society of America, 1963
- A Note on the Audibility of Intense Ultrasonic SoundThe Journal of the Acoustical Society of America, 1954
- Effect of Different Types of Electrodes in Electrophonic HearingThe Journal of the Acoustical Society of America, 1953
- Zur Theorie des Hörens bei der Schallaufnahme durch KnochenleitungAnnalen der Physik, 1932