One-third of dilated cardiomyopathy patients receiving a biventricular pacing-device do not respond to this form of therapy. Therefore, the utility of mechanical dyssynchrony by real-time 3D echocardiography (RT3DE) for predicting systolic response to biventricular pacing, of which maximal rate of pressure rise (d P /d tmax ) served as the gold-standard, was evaluated. Seventeen consecutive heart failure patients (aged 64 ± 10 years, 8 male, 6 ischaemic cardiomyopathy, mean QRS duration 136 ± 32 ms) underwent RT3DE and biventricular pacing. Post-processing software provided data of global left ventricular (LV) function and the systolic dyssynchrony index of 17 LV segments (SDI 17 , %) for mechanical dyssynchrony. During biventricular pacing, percentual change in d P /d tmax compared to the non-pacing mode, Δd P /d tmax was measured invasively with conductance catheters. LV ejection fraction was 31 ± 10%, SDI 17 was 10.2 ± 4.2% and percentual Δd P /d tmax during biventricular pacing was 14.5 ± 12.4. A significant correlation ( r = 0.729, P = 0.001) was found between SDI 17 and percentual Δd P /d tmax , and between QRS duration and percentual Δd P /d tmax ( r = 0.721, P = 0.001). The present study suggests that mechanical dyssynchrony measured by RT3DE shows a good correlation with invasively determined acute haemodynamic response to biventricular pacing in patients with symptomatic dilated cardiomyopathy. Future studies are needed to further define the clinical utility of RT3DE in identifying patients who are most likely to respond to cardiac resynchronization therapy.