Evaluation of PTW Seven29 for tomotherapy patient-specific quality assurance and comparison with ScandiDos Delta 4
Open Access
- 1 January 2012
- journal article
- research article
- Published by Medknow in Journal of Medical Physics
- Vol. 37 (2) , 72-80
Abstract
For routine quality assurance of helical tomotherapy plans, an alternative method, as opposed to the TomoTherapy suggested cylindrical solid water phantom with film and ionization chamber, is proposed using the PTW Seven29 2D-ARRAY inserted in a dedicated octagonal phantom, called Octavius. First, the sensitivity of the array to pitch was studied by varying the pitch during planning to 0.287, 0.433, 1.0, and 2.0. For each pitch selected, the dependence on field size was investigated by generating plans with field widths (FWs) of 1.06 cm, 2.49 cm, and 5.02 cm, for a total of 12 plans. Secondly, a total of 15 patient QA plans were delivered using helical tomotherapy with the Delta 4 and Seven29/Octavius for comparison. Using the clinical gamma criteria, 3% and 3 mm, all FW and pitch plans had a passing percentage of >90%. For patient QA plans, the average gamma pass percentage was 97.0% (94.4-99.8%) for the Delta 4 and 97.6% (92.5-100.0%) for the Seven29/Octavius. Both the Seven29/Octavius and Delta 4 performed to a high standard of measurement accuracy and had a 90% or greater gamma percent for all plans and were considered clinically acceptable.Keywords
This publication has 19 references indexed in Scilit:
- Recent advances in radiotherapyBMC Medicine, 2010
- IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119Medical Physics, 2009
- Longitudinal study using a diode phantom for helical tomotherapy IMRT QAMedical Physics, 2009
- An intercomparison between film dosimetry and diode matrix for IMRT quality assuranceMedical Physics, 2007
- D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancerPhysics in Medicine & Biology, 2005
- Characterization of a 2D ion chamber array for the verification of radiotherapy treatmentsPhysics in Medicine & Biology, 2005
- Verification of IMRT fields by film dosimetryMedical Physics, 2003
- Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluationMedical Physics, 2002
- Dose verification for patients undergoing IMRTMedical Dosimetry, 2001
- Dosimetric verification of a commercial inverse treatment planning systemPhysics in Medicine & Biology, 1999