Intraprocedural assessment of valve geometry during transcatheter mitral valve replacement by large field-of-view intravascular ultrasound: a case report
Open Access
- 20 December 2021
- journal article
- editorial
- Published by Oxford University Press (OUP) in European Heart Journal - Case Reports
- Vol. 5 (12) , ytab508
Abstract
A 59-year-old male 9 years post-surgical mitral valve replacement (29 mm Carpentier-Edwards PERIMOUNT Plus, Edwards Lifesciences, Irvine, USA) for severe mitral regurgitation, presented with significant degenerative bioprosthesis stenosis and severe insufficiency ( Figure 1A). Due to a EuroSCORE-II of 10.2%, valve-in-valve transcatheter mitral valve replacement (TMVR) was scheduled. Cardiac computed CT angiography (CCTA) revealed a small leaflet calcium deposit (yellow arrow) with inner ring diameters of 2.51 cm × 2.73 cm ( Figure 1B). A 10 MHz Vision PV 0.035 intravascular ultrasound (IVUS, Philips North America Corporation, Andover, USA) with a 60 mm imaging field tracking over a 0.035″ standard guidewire showed calcified commissure ( Figure 1C; white arrow indicates the transducer non-coaxial location) (yellow arrow) with thickened, poorly mobile, non-coapting leaflets (white arrows). 1–3 Coaptation gap measured 65.8 mm2 and inner ring 27.3 mm ( Figure 1D; asterisk indicates IVUS transducer). Angiographically successful SAPIEN3 deployment was verified with IVUS ( Figure 1E; white arrow indicates the transducer coaxial location). SAPIEN3 inflow (overlapping the bioprosthesis ring) was elliptical with outer frame diameters of 25.7 mm × 31.8 mm [eccentricity index (EI) of 1.23] and 97% expansion (% of nominal area; 643/660 mm2; Figure 1F). The more round mid-segment (27.8 mm × 30.1 mm; EI of 1.08) had 100% expansion (665/660 mm2). Angles between the neo-commissures differed (95° vs. 120° vs. 145°). Geometric orifice was elliptical at the valve inflow (405.5 mm2 in diastole; EI of 1.26) and more round at the mid-segment (343.9 mm2; EI of 1.12). The transoesophageal echocardiography (TOE) residual transvalvular max/mean diastolic gradient was 4.6/1.8 mmHg with normal leaflet function (3D enface view, Figure 1G). On Day 3, the transthoracic echo gradient increased to 15.7/10.1 mmHg despite normal leaflet function.Keywords
This publication has 3 references indexed in Scilit:
- Intravascular ultrasound for valve frame expansion and orifice area measurement as well as paravalvular leak assessment during transcatheter aortic valve replacementEuropean Heart Journal, 2021
- Intravascular ultrasound for transcatheter pulmonary valve replacementPolish Heart Journal, 2020
- Intravascular ultrasound online guidance during transcatheter valve replacement for native aortic stenosis or failed bioprosthesisPolish Heart Journal, 2020