The limitations of113Inmfor plasma volume measurement
Author:
- R. WoottondRWR. Woottond
- 1 May 1976
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 49 (581) , 427-429
Abstract
Previous authors have claimed that 113InmCl3 is a valid tracer for plasma volume measurement. However, careful measurements in haematologically normal volunteers and in rabbits, using 125I-labelled human serum albumin as a reference tracer, show that the results obtained with 113Inm-chloride, -citrate or -transferrin consistently over-estimate the plasma volume and are probably no better than can be predicted from the height and weight of the subject.Keywords
This publication has 1 reference indexed in Scilit:
- Measurement of plasma volume using99Tcm, and113Inmlabelled proteinsBritish Journal of Radiology, 1969