In vitro bench testing using patient-specific 3D models for percutaneous pulmonary valve implantation with Venus P-valve
Open Access
- 22 August 2023
- journal article
- research article
- Published by Wolters Kluwer Health in Chinese Medical Journal
- Vol. 137 (8) , 990-996
Abstract
Background: Due to the wide variety of morphology, size, and dynamics, selecting an optimal valve size and location poses great difficulty in percutaneous pulmonary valve implantation (PPVI). This study aimed to report our experience with in vitro bench testing using patient-specific three-dimensional (3D)-printed models for planning PPVI with the Venus P-valve. Methods: Patient-specific 3D soft models were generated using PolyJet printing with a compliant synthetic material in 15 patients scheduled to undergo PPVI between July 2018 and July 2020 in Central China Fuwai Hospital of Zhengzhou University. Results: 3D model bench testing altered treatment strategy in all patients (100%). One patient was referred for surgery because testing revealed that even the largest Venus P-valve would not anchor properly. In the remaining 14 patients, valve size and/or implantation location was altered to avoid valve migration and/or compression coronary artery. In four patients, it was decided to change the point anchoring because of inverted cone-shaped right ventricular outflow tract (RVOT) (n = 2) or risk of compression coronary artery (n = 2). Concerning sizing, we found that an oversize of 2–5 mm suffices. Anchoring of the valve was dictated by the flaring of the in- and outflow portion in the pulmonary artery. PPVI was successful in all 14 patients (absence of valve migration, no coronary compression, and none-to-mild residual pulmonary regurgitation [PR]). The diameter of the Venus P-valve in the 3D simulation group was significantly smaller than that of the conventional planning group (36 [2] vs. 32 [4], Z = -3.77, P In vitro testing indicated no need to oversize the Venus P-valve to the degree recommended by the balloon-sizing technique, as 2–5 mm sufficed.Keywords
This publication has 24 references indexed in Scilit:
- Early European experience with the Venus P-valve®: filling the gap in percutaneous pulmonary valve implantationEuroIntervention, 2016
- Patient-Specific Computer Modeling to Predict Aortic Regurgitation After Transcatheter Aortic Valve ReplacementJACC: Cardiovascular Interventions, 2016
- Patient-specific image-based computer simulation for theprediction of valve morphology and calcium displacement after TAVI with the Medtronic CoreValve and the Edwards SAPIEN valveEuroIntervention, 2016
- Percutaneous Pulmonary Valve ImplantationJACC, 2015
- Percutaneous pulmonary valve implantation with the Venus P-valve: clinical experience and early resultsCardiology in the Young, 2015
- Multi-modal imaging support in a staging percutaneous pulmonary valve implantationEuropean Heart Journal, 2015
- Early clinical experience with a novel self‐expanding percutaneous stent‐valve in the native right ventricular outflow tractCatheterization and Cardiovascular Interventions, 2014
- Rapid prototyping compliant arterial phantoms for in-vitro studies and device testingJournal of Cardiovascular Magnetic Resonance, 2013
- Virtual and real bench testing of a new percutaneous valve device: a case studyEuroIntervention, 2012
- Variations in Right Ventricular Outflow Tract Morphology Following Repair of Congenital Heart Disease: Implications for Percutaneous Pulmonary Valve ImplantationJournal of Cardiovascular Magnetic Resonance, 2007