The Variability of Applicator Position among High Dose Rate Intracavitary Brachytherapy Applications in Cervical Cancer Patients Treated with Ring & Tandem Applicators
Authors:
- 1 September 2007
- journal article
- research article
- Published by SAGE Publications in Tumori Journal
- Vol. 93 (5) , 432-438
Abstract
Objective: To evaluate the interindividual and intraindividual applicator position variability in high dose rate ring and tandem intracavitary brachytherapy applications in locally advanced cervical cancer. Patients and Methods: Eight patients with locally advanced cervical cancer formed the study population. They had been treated in Dokuz Eylül University Department of Radiation Oncology between the years 2000 and 2005 with high dose rate intracavitary brachytherapy using ring and tandem applicators. The 3-dimensional geometric variation of the applicator center in craniocaudal, mediolateral and anteroposterior directions was determined on the basis of bony reference points in 24 pairs of orthogonal films obtained in the conventional simulator. Then the following evaluations were performed: 1) the applicator position variability in all applications (interindividual variability), 2) the intraindividual applicator position variability relative to the first application, 3) the intraindividual applicator position variability relative to the average of three applications. Among the potential factors that might influence the reproducibility of ring and tandem applications, age, stage, the period between external radiotherapy and brachytherapy were evaluated by univariate analysis. Results: Standard deviation of interindividual applicator variability was 3.83 mm in craniocaudal, 0.39 mm in mediolateral and 2.86 mm in anteroposterior directions. The standard deviation of intraindividual variability relative to the first application was 1.91 mm in craniocaudal, 0.4 mm in mediolateral, and 4.26 mm in anteroposterior directions. The standard deviation of intraindividual variability relative to the average of three applications was 0.95 mm in craniocaudal, 1.86 mm in mediolateral, and 1.24 mm in anteroposterior directions. According to univariate analysis, no factor influenced applicator position variability. Conclusions: In order to extract definitive conclusions about factors that affect positional reproducibility of ring and tandem applicators, studies are needed that include larger numbers of patients.Keywords
This publication has 13 references indexed in Scilit:
- Positional Variability of a Tandem Applicator System in HDR Brachytherapy for Primary Treatment of Cervix CancerStrahlentherapie und Onkologie, 2004
- Variations of intracavitary applicator geometry during multiple HDR brachytherapy insertions in carcinoma cervix and its influence on reporting as per ICRU report 38Radiotherapy and Oncology, 2001
- Spatial Reproducibility of the Ring and Tandem High-Dose Rate Cervix ApplicatorInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1998
- Midline shield for radiation therapy of carcinoma of the uterine cervix: Should it be “midline” or “individualized”?Medical Dosimetry, 1997
- Changes in applicator position with fractionated high dose rate gynaecological brachytherapyRadiotherapy and Oncology, 1996
- Dose-volume histogram analysis of high dose rate intracavitary brachytherapy for uterine cervix cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1996
- Positional stability of sources during low-dose-rate brachytherapy for cervical carcinomaInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1993
- Patterns and sites of failure in cervix cancer treated in the U.S.A. in 1978International Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991
- Radiation alone in the treatment of cancer of the uterine cervix: Analysis of pelvic failure and dose response relationshipInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1989