A Novel Calibration-Free Method of Measurement of Oxygen Saturation in Arterial Blood
- 10 February 2009
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 58 (5) , 1699-1705
Abstract
In present-day pulse oximeters, oxygen saturation in arterial blood (SpO2) is computed by utilizing an empirical relationship extracted from a calibration curve. The calibration curve is obtained by curve-fitting data acquired from volunteers. A novel method of computation of SpO2 that does not require the use of a calibration curve is presented in this paper. Based on a model for the attenuation of light through skin, tissue, bone, and blood, suitable processing steps are identified so that the analytical expression derived for the estimation of SpO2 becomes free of not only patient but sensor-dependent parameters as well. The experimental results presented in this paper establish the efficacy of the proposed method.Keywords
This publication has 12 references indexed in Scilit:
- The measurement of oxygen saturation in arterial and venous bloodIEEE Instrumentation & Measurement Magazine, 2008
- A Novel Method of Measurement of Oxygen Saturation in Arterial Blood2008
- Extinction coefficients of hemoglobin for near-infrared spectroscopy of tissueIEEE Engineering in Medicine and Biology Magazine, 2005
- Pulse Oximetry Theory and Calibration for Low SaturationsIEEE Transactions on Biomedical Engineering, 2004
- Pulse oximetry: Principles and limitationsThe American Journal of Emergency Medicine, 1999
- Computer analysis system of blood oxygen saturation in an animal hypoxia model.Medical & Biological Engineering & Computing, 1998
- Wavelength selection for low-saturation pulse oximetryIEEE Transactions on Biomedical Engineering, 1997
- Design of Pulse Oximeters1997
- 8 Fetal pulse oximetryBailliere's Clinical Obstetrics and Gynaecology, 1996
- Signal processing methods for pulse oximetryComputers in Biology and Medicine, 1996