Interdevice Agreement of Keratometry, Astigmatism Vectors, and Ocular Biometry in Cataract Candidates: SS-OCT (Argos) vs. OLCI (Aladdin) vs. Scheimpflug–Placido (Sirius)
Open Access
- 3 March 2026
- journal article
- research article
- Published by MDPI AG in Bioengineering
- Vol. 13 (3) , 296
Abstract
Background and Objectives: Accurate anterior segment measurements are central to intraocular lens (IOL) power calculation and toric planning, yet different optical platforms may yield non-interchangeable values. This study compared keratometry, astigmatism metrics, and ocular biometry obtained with a swept-source OCT biometer (Argos), an optical low-coherence interferometry biometer (Aladdin), and a combined Scheimpflug–Placido topographer (Schwind Sirius). Methods: This is a retrospective observational study (January 2022–June 2024) including eyes undergoing uncomplicated cataract surgery. All eyes were measured in a single session by one examiner. Outcomes included K1, K2, cylinder, astigmatism axis (degrees; device-reported corneal cylinder axis, labeled “Powerful Angle” in the Sirius export), vector components (J0 and J45), and—where available—lens thickness (LT), axial length (AL), anterior chamber depth (ACD), white-to-white (WTW) distance, and central corneal thickness (CCT). Friedman tests assessed 3-device differences, and pairwise comparisons were evaluated using Wilcoxon signed-rank tests (paired data). Results: A total of 170 eyes (102 patients) were analyzed (mean age: 69.12 ± 10.26 years). Significant inter-device differences were detected for K1 (Argos: 43.45 ± 1.64 D; Aladdin: 43.41 ± 1.70 D; overall: p < 0.001; Argos vs. Aladdin: p = 0.019), K2 (Argos: 44.45 ± 1.67 D; Aladdin: 44.34 ± 1.71 D; overall and pairwise: p < 0.001), and cylinder (Argos: −0.83 ± 0.74 D, Aladdin: −0.77 ± 0.76 D; Sirius: −0.68 ± 0.75 D; overall: p < 0.001). “Powerful Angle” differed across devices (p = 0.003) but not between Argos and Aladdin (p = 0.512). J0 (p = 0.277) and J45 (p = 0.084) did not differ significantly. Argos reported higher ACD (3.19 ± 0.42 vs. 3.13 ± 0.41 mm, p < 0.001) and WTW (11.95 ± 0.42 vs. 11.65 ± 0.39 mm, p < 0.001) values than Aladdin. CCT was similar between Aladdin and Sirius (540.27 ± 33.44 vs. 540.47 ± 33.78 µm, p = 0.169). Conclusions: Several keratometric and biometric parameters differed significantly by device, indicating limited interchangeability—particularly relevant for toric and premium IOL planning—while vector astigmatism components and CCT showed better agreement.Keywords
All Related Versions
Funding Information
- Victor Babeș University of Medicine and Pharmacy in Timișoara
This publication has 23 references indexed in Scilit:
- Harmonizing NAD + metabolism with CeMOF nanozyme to orchestrate intraocular pressure modulation for glaucoma treatmentView, 2025
- Comparison of keratometric values and anterior segment parameters measured using Scheimpflug Sirius topography and Lenstar biometryInternational Ophthalmology, 2024
- Evaluation of three biometric devices: ocular parameters and calculated intraocular lens powerScientific Reports, 2022
- Comparison of ocular biometric measurements in patients with cataract using three swept-source optical coherence tomography devicesBMC Ophthalmology, 2021
- Comparison of biometry and intraocular lens power calculation performed by a new optical biometry device and a reference biometerJournal of Cataract & Refractive Surgery, 2017
- Contribution of posterior corneal astigmatism to total corneal astigmatismJournal of Cataract & Refractive Surgery, 2012
- Local anaesthesia for trans‐scleral cyclodiode laser procedures: surgeon and patient satisfaction with sub‐Tenon's and peribulbar anaesthesiaClinical & Experimental Ophthalmology, 2011
- Anterior chamber angle imaging with optical coherence tomographyEye, 2011
- Power vector analysis of the optical outcome of refractive surgeryJournal of Cataract & Refractive Surgery, 2001
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992