Automatic AMS Sample Combustion and CO2 Collection
- 1 January 2001
- journal article
- Published by Cambridge University Press (CUP) in Radiocarbon
- Vol. 43 (2A) , 293-298
Abstract
In Groningen, all organic samples for accelerator mass spectrometry (AMS) are combusted in an automatic Elemental Analyzer, coupled to an Isotope Ratio Mass Spectrometer and Cryogenic Trapping System. The Gas Chromatographic (GC) column, part of the Elemental Analyzer system, appeared to be the main cause for memory effects. Therefore we modified the Elemental Analyzer, such that the trapped CO2 no longer passed the GC column. Our system modification reduced the memory effect significantly, as shown by lower radiocarbon concentration values for anthracite backgrounds, and a much smaller spread in these values. Our modified system can perform up to 40 combustions unattended in about 6 hr.Keywords
This publication has 6 references indexed in Scilit:
- Status report: The Groningen AMS facilityNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2000
- The Groningen AMS tandetronNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1997
- First Results from the Groningen AMS FacilityRadiocarbon, 1995
- The Hvee 14C System at GroningenRadiocarbon, 1995
- A High Throughput 14C Accelerator Mass SpectrometerRadiocarbon, 1992
- The IAEA 14C Intercomparison Exercise 1990Radiocarbon, 1992