Picomolar‐scale compound‐specific isotope analyses
- 15 February 2018
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 32 (9) , 730-738
Abstract
Rationale: We report modifications to compound‐specific isotope analyses (CSIA) to enable high‐precision isotopic analyses of picomoles of carbon for intact organic molecules. This sample size is two orders of magnitude below the amounts required for commercial systems. The greatly enhanced sensitivity of this system expands molecular isotope studies and applications previously prohibited by low concentrations and small samples. Methods: We utilize the resolving power and low volumetric flow rates of narrow‐bore capillary gas chromatography to improve sample transfer efficiency while maintaining narrow peak widths. Post‐column peak broadening is minimized using a micro‐fluidic valve for solvent diversion, capillary combustion reactor, narrow‐bore capillary transfer lines, and cryogenic water trap. The mass spectrometer was fitted with collector amplifiers configured to 25 ms response times and a data logger board with firmware capable of rapid data acquisition. Carbon dioxide gas was introduced directly into the ion source to evaluate the dynamic range of the system and accuracy and precision of carbon isotope ratio (δ 13 C value) measurements. The accuracy and precision for combusted compounds were evaluated using a suite of n ‐alkanes. Results: For ≥30 pmol carbon introduced directly into the ion source, the mean difference between the measured and expected δ 13 C values is 0.03‰ (1 σ , n = 57) and the standard deviation of replicate measurements is 0.11‰ (1 σ ). The CO 2 peak widths generated by the exponential dilution flask were 250 ms and the peak widths produced by combusting n ‐alkanes were ca 500 ms, less than 25% the width of conventional gas chromatography peaks. For a mixture of 15 n ‐alkanes ( n ‐C 16 to n ‐C 30 ), the accuracy is 0.3‰ (1 σ ) and precision is 0.9‰ (1 σ ) for replicate δ 13 C measurements with 100 pmol carbon per compound on column. Conclusions: The pico‐CSIA method described here offers improved chromatographic resolution and reduces sample size requirements by two orders of magnitude. These advances significantly broaden the available analytical window for CSIA in research areas frequently hindered by sample size limitations, such as forensics, paleoclimate, astrobiology, and biochemistry.Keywords
Funding Information
- National Aeronautics and Space Administration (NX13AI80G)
This publication has 27 references indexed in Scilit:
- An improved and rapid 5A molecular sieve method for gas chromatography isotope ratio mass spectrometry of n-alkanes (C8–C30+)Organic Geochemistry, 2008
- ConFlo III - an interface for high precision δ13C and δ15N analysis with an extended dynamic rangeRapid Communications in Mass Spectrometry, 1999
- Hardware Modifications to an Isotope Ratio Mass Spectrometer Continuous-Flow Interface Yielding Improved Signal, Resolution, and MaintenanceAnalytical Chemistry, 1998
- Performance and Optimization of a Combustion Interface for Isotope Ratio Monitoring Gas Chromatography/Mass SpectrometryAnalytical Chemistry, 1995
- Factors Controlling Precision and Accuracy in Isotope-Ratio-Monitoring Mass SpectrometryAnalytical Chemistry, 1994
- Acquisition and processing of data for isotope-ratio-monitoring mass spectrometryOrganic Geochemistry, 1994
- Compound-specific isotopic analyses: A novel tool for reconstruction of ancient biogeochemical processesOrganic Geochemistry, 1990
- Evidence from carbon isotope measurements for diverse origins of sedimentary hydrocarbonsNature, 1990
- Gas chromatography—stable isotope ratio analysis at natural abundance levelsJournal of Mass Spectrometry, 1984
- Isotope-ratio-monitoring gas chromatography-mass spectrometryAnalytical Chemistry, 1978