Three-dimensional evaluation of the mitral valve area and commissural opening before and after percutaneous mitral commissurotomy in patients with mitral stenosis
Authors:
- David Messika-ZeitounDMDavid Messika-Zeitoun
- Eric BrochetEBEric Brochet
- Caroline HolminCHCaroline Holmin
- David RosenbaumDRDavid Rosenbaum
- Bertrand CormierBCBertrand Cormier
- Jean-Michel SerfatyJSJean-Michel Serfaty
- Bernard IungBIBernard Iung
- Alec VahanianAVAlec Vahanian
- 13 June 2006
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 28 (1) , 72-79
Abstract
Aims Management of patients with mitral stenosis (MS) relies on accurate evaluation of the mitral valve area (MVA). Planimetry (MVA2D) is considered as the reference method but must be performed at the tips of the leaflets with the correct plane orientation and therefore requires experienced operators. Real-time three-dimensional echocardiography (RT3DE) may overcome this limitation but its usefulness for experienced when compared with less experienced operators has not been evaluated. In addition, superiority of RT3DE for the evaluation of commissural splitting after percutaneous mitral commissurotomy (PMC) is unknown. Methods and results 60 patients were prospectively evaluated by 2D and RT3DE before and after PMC by experienced operators. Before PMC, MVA3D was slightly higher than MVA2D (1.15 ± 0.25 vs. 1.06 ± 0.22 cm2, P = 0.0001) but correlation between methods was excellent ( r = 0.73, P < 0.0001), mean difference was small (0.09 ± 0.18 cm2) and clinically meaningless (three patients misclassified, two of whom had borderline MS severity). After PMC, MVA3D did not differ from and correlated well with MVA2D (1.87 ± 0.37 vs. 1.85 ± 0.32 cm2, P = 0.36; r = 0.76, P < 0.0001; mean difference 0.03 ± 0.24 cm2). Twenty-five additional patients were also evaluated both by an experienced and a less experienced operators. Bland–Altman analysis showed the better agreement between MVA3D measured by the less experienced operator and MVA2D measured by the experienced operator than between MVA2D measured by the less experienced and the experienced operators (mean difference 0.03 ± 0.34 vs. − 0.13 ± 0.46 cm2, P = 0.03). When compared with RT3DE, 2DE underestimated the degree of commissural opening in 33% of patients and agreement between methods was weak ( κ = 0.41). Conclusion RT3DE provides accurate MVA measurements similar to 2D planimetry performed by experienced operators. Thus, it does not provide a real advantage for experienced operators, whereas it seems particularly helpful for less experienced operators. In addition, RT3DE improves the description of valvular anatomy.Keywords
This publication has 16 references indexed in Scilit:
- Comparison of Accuracy of Mitral Valve Area in Mitral Stenosis by Real-Time, Three-Dimensional Echocardiography Versus Two-Dimensional Echocardiography Versus Doppler Pressure Half-TimeThe American Journal of Cardiology, 2005
- Non-invasive assessment of mitral valve area during percutaneous balloon mitral valvuloplasty: role of real-time 3D echocardiographyEuropean Heart Journal, 2004
- Real-time three-dimensional echocardiography for rheumatic mitral valve stenosis evaluationJACC, 2004
- Additional value of three-dimensional transesophageal echocardiography for patients with mitral valve stenosis undergoing balloon valvuloplastyJournal of the American Society of Echocardiography, 2003
- A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on Valvular Heart DiseaseEuropean Heart Journal, 2003
- Recommendations for evaluation of the severity of native valvular regurgitation with two-dimensional and doppler echocardiographyJournal of the American Society of Echocardiography, 2003
- Utility of three-dimensional echocardiography during balloon mitral valvuloplastyJACC, 1998
- Variability between methods of calculating mitral valve area: Simultaneous Doppler echocardiographic and cardiac catheterization studies conducted before and after percutaneous mitral valvuloplastyJournal of the American Society of Echocardiography, 1996
- Three-dimensional echocardiography of normal and pathologic mitral valve: A comparison with two-dimensional transesophageal echocardiographyJACC, 1996
- Revision of Gorlin constant for calculation of mitral valve area from left heart pressures.Heart, 1973