Abstract
A novel non-enzymatic electrochemical sensor was constructed to deal with the problems related to serious coronary heart diseases for highly selective and ultrasensitive detection of cholesterol. The NiVP/Pi based sensor exhibits ultra-high sensitivity of 5510.18 μA μM−1 cm−2 and other at 36.8 μA μM−1 cm−2 from 1 nM to 10 μM and 100 μM to 10 mM range, respectively and ultra-low detection limit of 1 aM, along with superior selectivity towards various interferents like ascorbic acid, glucose and uric acid. Besides, novel NiVP/Pi based flexible sensor coated over Whatman filter paper was developed and displayed superior sensitivity, even with human blood serum sample at physiological pH in physiologically range.
Funding Information
  • Anusandhan National Research Foundation (CRG/2018/004478)
  • Department of Science and Technology, Ministry of Science and Technology (SR/NM/ NS-1376/2014)
  • Council of Scientific and Industrial Research, India (02(0436)/21/EMRII)

This publication has 15 references indexed in Scilit: