An Analytical Calculation of Frictional and Bending Moments at the Head-Neck Interface of Hip Joint Implants during Different Physiological Activities
Open Access
Abstract
This study predicts the frictional moments at the head-cup interface and frictional torques and bending moments acting on the head-neck interface of a modular total hip replacement across a range of activities of daily living. The predicted moment and torque profiles are based on the kinematics of four patients and the implant characteristics of a metal-on-metal implant. Depending on the body weight and type of activity, the moments and torques had significant variations in both magnitude and direction over the activity cycles. For the nine investigated activities, the maximum magnitude of the frictional moment ranged from 2.6 to 7.1 Nm. The maximum magnitude of the torque acting on the head-neck interface ranged from 2.3 to 5.7 Nm. The bending moment acting on the head-neck interface varied from 7 to 21.6 Nm. One-leg-standing had the widest range of frictional torque on the head-neck interface (11 Nm) while normal walking had the smallest range (6.1 Nm). The widest range, together with the maximum magnitude of torque, bending moment, and frictional moment, occurred during one-leg-standing of the lightest patient. Most of the simulated activities resulted in frictional torques that were near the previously reported oxide layer depassivation threshold torque. The predicted bending moments were also found at a level believed to contribute to the oxide layer depassivation. The calculated magnitudes and directions of the moments, applied directly to the head-neck taper junction, provide realistic mechanical loading data for in vitro and computational studies on the mechanical behaviour and multi-axial fretting at the head-neck interface.Keywords
This publication has 32 references indexed in Scilit:
- Friction in Total Hip Joint Prosthesis Measured In Vivo during WalkingPLOS ONE, 2013
- Fretting corrosion of CoCrMo and Ti6Al4V interfacesBiomaterials, 2012
- Progress in standardization of fretting fatigue terminology and testingTribology International, 2011
- Fretting crevice corrosion of stainless steel stem–CoCr femoral head connections: Comparisons of materials, initial moisture, and offset lengthJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- Friction moments of large metal-on-metal hip joint bearings and other modern designsMedical Engineering & Physics, 2008
- Neck fracture of a cementless forged titanium alloy femoral stem following total hip arthroplasty: a case report and review of the literatureJournal of Medical Case Reports, 2007
- Transient elastohydrodynamic lubrication analysis of metal-on-metal hip implant under simulated walking conditionsJournal of Biomechanics, 2006
- In vitro corrosion testing of modular hip tapersJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2002
- Heat shock‐induced necrosis and apoptosis in osteoblastsJournal of Orthopaedic Research, 1999
- In vivo records of hip loads using a femoral implant with telemetric output (a prelimary report)Journal of Biomedical Engineering, 1979