Acute Atrial Pacing Threshold Elevation in an Auricular Leadless Pacemaker During Monopolar Pulsed Field Ablation of the Right Inferior Pulmonary Vein: A First‐in‐Human Case
- 14 January 2026
- journal article
- case report
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 37 (3) , 658-662
Abstract
Introduction: Leadless atrial pacemakers are increasingly used, yet interactions with pulsed field ablation (PFA) remain poorly described. Methods and Results: We report the first documented case of transient loss of capture and pacing threshold elevation in an AVEIR leadless pacemaker temporally associated with monopolar PFA during right inferior pulmonary vein ablation. Threshold increased from 0.75 V @ 0.2 ms to 1.25 V @ 0.4 ms, recovering within 24 h. Conclusion: Monopolar PFA can transiently interfere with atrial leadless pacing, emphasizing the need for device monitoring during ablation.Keywords
This publication has 3 references indexed in Scilit:
- Cardiac implantable electronic device software reset and damage after pulsed field ablation using a monopolar lattice-tip catheterHeart Rhythm, 2025
- Elevated pacing thresholds during atrial leadless pacemaker implantation: Accept or reposition?Heart Rhythm, 2025
- Safety of pulsed field ablation using a variable loop circular catheter in a patient with a cardiac implantable electronic deviceHeartRhythm Case Reports, 2024