An HS-MRM Assay for the Quantification of Host-cell Proteins in Protein Biopharmaceuticals by Liquid Chromatography Ion Mobility QTOF Mass Spectrometry
Open Access
- 17 April 2018
- journal article
- research article
- Published by MyJove Corporation in Journal of Visualized Experiments
- No. 134,p. e55325
Abstract
The analysis of low-level (1-100 ppm) protein impurities (e.g., host-cell proteins (HCPs)) in protein biotherapeutics is a challenging assay requiring high sensitivity and a wide dynamic range. Mass spectrometry-based quantification assays for proteins typically involve protein digestion followed by the selective reaction monitoring/multiple reaction monitoring (SRM/MRM) quantification of peptides using a low-resolution (Rs ~1,000) tandem quadrupole mass spectrometer. One of the limitations of this approach is the interference phenomenon observed when the peptide of interest has the "same" precursor and fragment mass (in terms of m/z values) as other co-eluting peptides present in the sample (within a 1-Da window). To avoid this phenomenon, we propose an alternative mass spectrometric approach, a high selectivity (HS) MRM assay that combines the ion mobility separation of peptide precursors with the high-resolution (Rs ~30,000) MS detection of peptide fragments. We explored the capabilities of this approach to quantify low-abundance peptide standards spiked in a monoclonal antibody (mAb) digest and demonstrated that it has the sensitivity and dynamic range (at least 3 orders of magnitude) typically achieved in HCP analysis. All six peptide standards were detected at concentrations as low as 0.1 nM (1 femtomole loaded on a 2.1-mm ID chromatographic column) in the presence of a high-abundance peptide background (2 µg of a mAb digest loaded on-column). When considering the MW of rabbit phosphorylase (97.2 kDa), from which the spiked peptides were derived, the LOQ of this assay is lower than 50 ppm. Relative standard deviations (RSD) of peak areas (n = 4 replicates) were less than 15% across the entire concentration range investigated (0.1-100 nM or 1-1,000 ppm) in this study.Keywords
This publication has 23 references indexed in Scilit:
- Selected Reaction Monitoring Mass Spectrometry for Absolute Protein QuantificationJournal of Visualized Experiments, 2015
- Quantitative Host Cell Protein Analysis Using Two Dimensional Data Independent LC–MSEAnalytical Chemistry, 2015
- Toward the complete characterization of host cell proteins in biotherapeutics via affinity depletions, LC-MS/MS, and multivariate analysismAbs, 2015
- Comprehensive tracking of host cell proteins during monoclonal antibody purifications using mass spectrometrymAbs, 2014
- Detection and correction of interference in SRM analysisMethods, 2013
- Identification and quantification of host cell protein impurities in biotherapeutics using mass spectrometryAnalytical Biochemistry, 2012
- A Computational Tool to Detect and Avoid Redundancy in Selected Reaction MonitoringMolecular & Cellular Proteomics, 2012
- Quantification of Proteins Using Peptide Immunoaffinity Enrichment Coupled with Mass SpectrometryJournal of Visualized Experiments, 2011
- Automated Detection of Inaccurate and Imprecise Transitions in Peptide Quantification by Multiple Reaction Monitoring Mass SpectrometryClinical Chemistry, 2010
- Full Dynamic Range Proteome Analysis of S. cerevisiae by Targeted ProteomicsCell, 2009