Air Bubbles Pass the Security System of the Dialysis Device Without Alarming
- 12 February 2007
- journal article
- Published by Wiley in Artificial Organs
- Vol. 31 (2) , 132-139
Abstract
During hemodialysis microembolic findings have been noted after the venous chamber (subclavian vein). The aim of this study was to evaluate if air could pass the venous chamber and, if so, if it passes the safety-system detector for air-infusion without triggering an alarm. Various in vitro dialysis settings were performed using regular dialysis devices. A dextran fluid was used instead of blood to avoid the risk of development of emboli. Optical visualization as well as recirculation and collection of eventual air into an intermediate bag were investigated. In addition, a specifically designed ultrasound monitor was placed after the venous air trap to measure the presence of eventual microbubbles. Speed of dialysis fluid was changed, as was the level of the fluid in the air trap. Thereby a fluid level was considered "high" if it was close to the top of the air trap and "low" if it was around the mid part of the air trap. By optical vision microbubbles were seen at the bottom of the air trap and could pass the air trap towards the venous line without alarming. During recirculation several mL of air were collected in an intermediate bag after the venous line. Ultrasound monitoring exhibited the presence of microbubbles of the size of approximately 5 microm upwards passing to the venous line in all runs performed. Amount of bubbles differed between devices and in general an increased fluid speed correlated significantly with the increased counts of microbubbles/min. No alarming of the detector occurred. A more concentrated fluid allowed higher counts/min when flow was increased to 600 mL/min. Data revealed that air passes the safety-sensor in the air trap without alarming. The presence of air increased in general with fluid speed and a lower fluid level in the air trap. Differences were present between devices. If this affects the patients has to be elucidated.Keywords
This publication has 11 references indexed in Scilit:
- The Sensor in the Venous Chamber Does Not Prevent Passage of Air Bubbles During HemodialysisArtificial Organs, 2007
- MicrobubblesChest, 2005
- Overproduction of reactive oxygen species in end‐stage renal disease patients: A potential component of hemodialysis‐associated inflammationHemodialysis International, 2005
- Hemodialysis machines and monitors2004
- Reduction of circulating microemboli in the subclavian vein of patients undergoing haemodialysis using pre-filled instead of dry dialysersNephrology Dialysis Transplantation, 2003
- Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodiafiltration using pulsed Doppler ultrasoundNephrology Dialysis Transplantation, 2002
- Identification of microemboli during haemodialysis using Doppler ultrasoundNephrology Dialysis Transplantation, 2000
- Activation of FXII during haemodialysisScandinavian Journal of Clinical and Laboratory Investigation, 1996
- Gas Bubble Detection in Fluid Lines by Means of Pulsed Doppler UltrasoundScandinavian Journal of Thoracic and Cardiovascular Surgery, 1985
- Doppler Ultrasonic Detection of Particulate Release During Hemodialysis with Cellulose Hollow-Fiber and Sorbent Suspension Reciprocating DialyzersArtificial Organs, 1984